二维斑点追踪超声心动图分析胎儿心功能中的可重复性挑战:来自FetalHQ的见解

Reproducibility Challenges in Fetal Cardiac Function Analysis with 2D Speckle-Tracking Echocardiography: Insights from FetalHQ.

作者信息

Mlodawski Jakub, Zmelonek-Znamirowska Anna, Pawlik Lukasz, Mlodawska Marta, Swiercz Grzegorz

机构信息

Collegium Medicum, Jan Kochanowski University in Kielce, 25-516 Kielce, Poland.

Clinic of Obstetrics and Gynecology, Provincial Combined Hospital in Kielce, 25-736 Kielce, Poland.

出版信息

J Clin Med. 2025 May 9;14(10):3301. doi: 10.3390/jcm14103301.

Abstract

: Functional assessment of the fetal heart remains a significant challenge in contemporary perinatology due to the absence of a universally accepted gold standard for such evaluations. The aim of this study was to evaluate the reproducibility of parameters derived from two-dimensional speckle-tracking echocardiography (2D STE) using the FetalHQ software. : We enrolled 87 pregnant women between 19 and 23 weeks of gestation who were undergoing mid-trimester screening at the Provincial Hospital Complex in Kielce. Two independent operators acquired 5 s cine-loops of four-chamber views (4CVs) according to a standardized protocol. Reproducibility was assessed by examining intra- and interobserver variability using the intraclass correlation coefficient (ICC) for several cardiac parameters, including the global sphericity index (GSI), global longitudinal strain (GLS), stroke volume (SV), and fractional area change (FAC). : Reproducibility varied substantially across the assessed parameters. The highest intraobserver reproducibility was observed for the 4CV GSI (ICC > 0.9). Moderate intraobserver reproducibility (ICCs ranging from 0.5 to 0.75) was noted for left ventricular (LV) parameters, such as end-diastolic area, end-systolic area, end-diastolic volume, and end-systolic volume. Interobserver variability demonstrated higher ICC values, with excellent reproducibility (ICC > 0.9) for the 4CV GSI and LV volume measurements. Good reproducibility (ICCs between 0.75 and 0.9) was observed for specific left ventricular segmental strain indices, whereas other parameters showed moderate to poor reproducibility (ICC < 0.5). : Two-dimensional speckle-tracking echocardiography (2D STE) using FetalHQ exhibits variable reproducibility, which is influenced by the choice of parameters, operator experience, and technical factors. This method holds potential for fetal cardiac assessment; however, additional research is required to improve its predictive accuracy and streamline the evaluation process for routine clinical application.

摘要

由于缺乏此类评估的普遍接受的金标准,胎儿心脏功能评估在当代围产医学中仍然是一项重大挑战。本研究的目的是使用FetalHQ软件评估二维斑点追踪超声心动图(2D STE)得出的参数的可重复性。

我们招募了87名妊娠19至23周的孕妇,她们正在凯尔采省综合医院进行孕中期筛查。两名独立操作人员根据标准化方案获取了四腔心视图(4CV)的5秒动态环。通过使用组内相关系数(ICC)检查观察者内和观察者间变异性,对包括整体球形指数(GSI)、整体纵向应变(GLS)、每搏输出量(SV)和面积变化分数(FAC)在内的几个心脏参数的可重复性进行了评估。

在所评估的参数中,可重复性差异很大。4CV GSI的观察者内可重复性最高(ICC>0.9)。左心室(LV)参数,如舒张末期面积、收缩末期面积、舒张末期容积和收缩末期容积,观察者内可重复性中等(ICC范围为0.5至0.75)。观察者间变异性显示出更高的ICC值,4CV GSI和左心室容积测量具有出色的可重复性(ICC>0.9)。特定左心室节段应变指数具有良好的可重复性(ICC在0.75至0.9之间),而其他参数显示出中等至较差的可重复性(ICC<0.5)。

使用FetalHQ的二维斑点追踪超声心动图(2D STE)表现出可变的可重复性,这受到参数选择、操作人员经验和技术因素的影响。这种方法在胎儿心脏评估方面具有潜力;然而,需要进一步研究以提高其预测准确性并简化常规临床应用的评估过程。