Brunner J, Seidel D
Z Urol Nephrol. 1985 Jul;78(7):381-6.
In simultaneous experiments the haemolysis in the double-needle dialysis was investigated in comparison to the single-needle technique. As SN-technique the 2-clamp technique with permanently going and start-stop function of the blood pump, respectively, was used. Another technique without arterial and with retarded venous clamp was tested. Furthermore, the pressure limits in the SND were varied. A slight increase of the haemolysis in the SND in comparison to the DND was stated. In the different SN-techniques no essential differences in the haemolysis could be proved, whereas an increase of the pressure limits led to a clear increase of the haemolysis.
在同步实验中,对双针透析与单针技术相比的溶血情况进行了研究。作为单针技术,分别使用了具有血泵持续运转和启停功能的双夹技术。还测试了另一种无动脉夹且静脉夹延迟的技术。此外,还改变了双针透析中的压力限制。结果表明,与双针透析相比,单针透析中的溶血略有增加。在不同的单针技术中,未证实溶血有本质差异,而压力限制的增加导致溶血明显增加。