Bindhu Sree Ch, Koka Pradeep, Sri I Kavya, Alex Jisa A, Rani K Deepa, Chander V Bhanu
Department of Oral and Maxillofacial Prosthodontics, Malla Reddy Dental College for Women, Hyderabad, Telangana, India.
Department of Oral and Maxillofacial Prosthodontics, Malla Reddy Institute of Dental Sciences, Hyderabad, Telangana, India.
J Pharm Bioallied Sci. 2025 May;17(Suppl 1):S789-S791. doi: 10.4103/jpbs.jpbs_615_25. Epub 2025 Apr 12.
This study aims to evaluate and compare the trueness of four commercially available digital scanners.
A prepared lower first molar in a typhodont model, featuring dimple markers at line angles, served as the reference. Caliper measurements of dimple distances provided the reference dataset. Four desktop scanners each performed 10 scans of the model. EXOCAD software was used to process the digital scans, generating study datasets. Scanner datasets were then compared to the caliper reference data to evaluate trueness.
Repeated measures analysis of variance (ANOVA) determined significant differences ( < 0.05) in buccolingual and mesiodistal trueness between four scanners. Descriptive statistics revealed the UP3D scanner was most accurate in the buccolingual direction, while both MEDIT and UP3D were most accurate mesiodistally.
The Dentium rainbow scanner which was the only white light scanner in this study, showed less trueness than other types of scanners. The UP3D scanner which works on structured light principle had higher trueness.
本研究旨在评估和比较四种市售数字扫描仪的准确性。
在一个具有线角酒窝标记的牙齿模型中制备一颗下颌第一磨牙作为参考。用卡尺测量酒窝距离提供参考数据集。四台台式扫描仪分别对该模型进行10次扫描。使用EXOCAD软件处理数字扫描,生成研究数据集。然后将扫描仪数据集与卡尺参考数据进行比较,以评估准确性。
重复测量方差分析(ANOVA)确定四种扫描仪在颊舌向和近远中向准确性方面存在显著差异(<0.05)。描述性统计显示,UP3D扫描仪在颊舌向最准确,而MEDIT和UP3D在近远中向最准确。
本研究中唯一的白光扫描仪Dentium彩虹扫描仪,其准确性低于其他类型的扫描仪。基于结构光原理工作的UP3D扫描仪具有更高的准确性。