Son Keunbada, Lee Kyu-Bok
Department of Dental Science, Graduate School, Kyungpook National University, Daegu 41940, Korea.
Advanced Dental Device Development Institute, Kyungpook National University, Daegu 41940, Korea.
Materials (Basel). 2020 Apr 9;13(7):1744. doi: 10.3390/ma13071744.
The purpose of this study was to evaluate the accuracy of dental three-dimensional (3D) scanners according to the types of teeth. A computer-aided design (CAD) reference model (CRM) was obtained by scanning the reference typodont model using a high-precision industrial scanner (Solutionix C500, MEDIT). In addition, a CAD test model (CTM) was obtained using seven types of dental 3D scanners (desktop scanners (E1 and DOF Freedom HD) and intraoral scanners (CS3500, CS3600, Trios2, Trios3, and i500)). The 3D inspection software (Geomagic control X, 3DSystems) was used to segment the CRM according to the types of teeth and to superimpose the CTM based on the segmented teeth. The 3D accuracy of the scanner was then analyzed according to the types of teeth. One-way analysis of variance (ANOVA) was used to compare the differences according to the types of teeth in statistical analysis, and the Tukey HSD test was used for post hoc testing (α = 0.05). Both desktop and intraoral scanners showed significant differences in accuracy according to the types of teeth ( < 0.001), and the accuracy of intraoral scanners tended to get worse from anterior to posterior. Therefore, when scanning a complete arch using an intraoral scanner, the clinician should consider the tendency for the accuracy to decrease from anterior to posterior.
本研究的目的是根据牙齿类型评估牙科三维(3D)扫描仪的准确性。通过使用高精度工业扫描仪(Solutionix C500,MEDIT)扫描参考全口义齿模型获得计算机辅助设计(CAD)参考模型(CRM)。此外,使用七种类型的牙科3D扫描仪(台式扫描仪(E1和DOF Freedom HD)和口内扫描仪(CS3500、CS3600、Trios2、Trios3和i500))获得CAD测试模型(CTM)。使用3D检测软件(Geomagic control X,3DSystems)根据牙齿类型对CRM进行分割,并基于分割后的牙齿对CTM进行叠加。然后根据牙齿类型分析扫描仪的3D准确性。在统计分析中,使用单因素方差分析(ANOVA)比较根据牙齿类型的差异,并使用Tukey HSD检验进行事后检验(α = 0.05)。台式扫描仪和口内扫描仪在准确性方面均根据牙齿类型显示出显著差异(<0.001),并且口内扫描仪的准确性从前向后趋于变差。因此,当使用口内扫描仪扫描完整牙弓时,临床医生应考虑准确性从前向后降低的趋势。