Fuentealba Melani, Vallejos Carolina, Díez Sergio, Santibáñez Mauricio
Departamento de Cs. Físicas, Universidad de La Frontera, Temuco 4811230, Chile.
Laboratorio de Radiaciones Ionizantes, Universidad de La Frontera, Temuco 4811230, Chile.
Gels. 2025 May 30;11(6):416. doi: 10.3390/gels11060416.
This study evaluates the impact of gadolinium-based contrast agents (Omniscan, Dotarem, and Gadovist) on the performance of PAGAT gel dosimeters using spectrophotometric analysis. Dosimeters were infused with gadolinium at concentrations ranging from 0 to 40 mg/mL and irradiated with a 6 MV photon beam over a dose range of 0-15 Gy. Regarding dosimeter behavior, Dotarem exhibited an enhancement in optical density prior to irradiation due to polymerization reactions between the dosimeter and the contrast agent starting at 10 mg/mL, which compromised optical readings above 20 mg/mL. Omniscan consistently showed 37.7% lower sensitivity than standard PAGAT across all concentrations and dose levels. Conversely, Gadovist enhanced sensitivity by up to 50% at 20 mg/mL, with additional gains at higher concentrations, although accompanied by saturation at lower dose levels. Radiological analysis showed that all tested concentrations maintained mass energy-absorption coefficient differences below 1% and water-equivalence in effective atomic number within 5% at 6 MV. These findings underscore the importance of selecting an appropriate contrast agent to enhance gel dosimeter sensitivity, particularly in low-dose regions where measurement uncertainty increases. Additionally, gadolinium-infused PAGAT gels show strong potential for assessing dose enhancement phenomena. Their sensitivity, threshold behavior, and radiological properties suggest they may be suitable for applications in dose enhancement dosimetry as well as conventional clinical settings.
本研究使用分光光度分析评估了钆基造影剂(欧乃影、多它灵和钆喷酸葡胺)对PAGAT凝胶剂量计性能的影响。剂量计中注入浓度范围为0至40 mg/mL的钆,并在0 - 15 Gy的剂量范围内用6 MV光子束进行照射。关于剂量计行为,多它灵在照射前由于剂量计与造影剂之间从10 mg/mL开始的聚合反应而使光密度增强,这损害了高于20 mg/mL的光学读数。在所有浓度和剂量水平下,欧乃影始终比标准PAGAT显示出低37.7%的灵敏度。相反,钆喷酸葡胺在20 mg/mL时灵敏度提高了50%,在更高浓度时还有额外增加,尽管在较低剂量水平时伴有饱和度。放射学分析表明,所有测试浓度在6 MV时的质量能量吸收系数差异保持在1%以下,有效原子序数的水等效性在5%以内。这些发现强调了选择合适的造影剂以提高凝胶剂量计灵敏度的重要性,特别是在测量不确定性增加的低剂量区域。此外,注入钆的PAGAT凝胶在评估剂量增强现象方面显示出强大潜力。它们的灵敏度、阈值行为和放射学特性表明它们可能适用于剂量增强剂量测定以及传统临床环境中的应用。