Direction and rate of socket wear in Charnley low-friction arthroplasty.

作者信息

Wroblewski B M

出版信息

J Bone Joint Surg Br. 1985 Nov;67(5):757-61. doi: 10.1302/0301-620X.67B5.4055876.

DOI:10.1302/0301-620X.67B5.4055876
PMID:4055876
Abstract

Twenty-two high-density polyethylene sockets from Charnley low-friction arthroplasties have been studied. Acrylic casts and shadowgraph techniques were used to measure both the real and radiographic rates of wear; these rates showed a significant correlation. In the sagittal plane, nine of the sockets had worn lateral to a line drawn vertically from the centre of curvature of the socket, 12 had worn medial to that line and only one was worn exactly in the line. In the coronal plane, nine sockets had worn in front of a similar vertical line, two behind that line and 11 had worn exactly in the line. Evidence of impingement of the neck of the stem onto the rim of the socket was found in 14 patients; this is considered to be one of the causes of socket loosening. The obvious solution is to reduce the diameter of the neck of the stem from 12.5 mm to 10 mm; provided that it is made of cold-formed, high nitrogen-content stainless steel, this narrower neck is strong enough not to fracture.

摘要

相似文献

1
Direction and rate of socket wear in Charnley low-friction arthroplasty.
J Bone Joint Surg Br. 1985 Nov;67(5):757-61. doi: 10.1302/0301-620X.67B5.4055876.
2
External wear of the polyethylene socket in cemented total hip arthroplasty.骨水泥型全髋关节置换术中聚乙烯髋臼杯的外磨损
J Bone Joint Surg Br. 1987 Jan;69(1):61-3. doi: 10.1302/0301-620X.69B1.3818735.
3
Observations on the direction of wear in Charnley sockets retrieved at revision.对翻修时取出的Charnley髋臼杯磨损方向的观察。
J Bone Joint Surg Br. 1998 Nov;80(6):1067-72. doi: 10.1302/0301-620x.80b6.7647.
4
Effect of activity levels on polyethylene wear in Charnley low-friction arthroplasty.活动水平对Charnley低摩擦人工关节置换术中聚乙烯磨损的影响。
J R Coll Surg Edinb. 1999 Jun;44(3):193-6.
5
Loosening of the socket in a 35-mm snap-fit prosthesis and the Charnley hip prosthesis. A roentgenographic evaluation of 321 cases operated upon because of osteoarthritis.35毫米卡扣式假体和Charnley髋关节假体中髋臼的松动。对321例因骨关节炎接受手术的病例进行的X线评估。
Clin Orthop Relat Res. 1988 Mar(228):63-8.
6
Wear of the cup in the Charnley LFA in the young patient.
J Bone Joint Surg Br. 2004 May;86(4):498-503.
7
Wear of the acetabular socket. Comparison of polyacetal and polyethylene.髋臼臼杯的磨损。聚缩醛与聚乙烯的比较。
Acta Orthop Scand. 1986 Jun;57(3):193-6. doi: 10.3109/17453678608994373.
8
Wear in uncemented porous and cemented polyethylene sockets: a randomised, radiostereometric study.非骨水泥型多孔聚乙烯髋臼杯与骨水泥型聚乙烯髋臼杯的磨损情况:一项随机放射立体测量学研究
J Bone Joint Surg Br. 1998 Mar;80(2):345-50. doi: 10.1302/0301-620x.80b2.8032.
9
Factors affecting aseptic failure of fixation after primary Charnley total hip arthroplasty. Multivariate survival analysis.初次Charnley全髋关节置换术后影响固定无菌性失败的因素。多变量生存分析。
J Bone Joint Surg Am. 1997 Nov;79(11):1618-27. doi: 10.2106/00004623-199711000-00002.
10
The correlation between the roentgenographic appearance and operative findings at the bone-cement junction of the socket in Charnley low friction arthroplasties.
Clin Orthop Relat Res. 1988 Mar(228):105-9.

引用本文的文献

1
Long-term results of Charnley low friction arthroplasty in patients younger than 40.
Eur J Orthop Surg Traumatol. 1996 Dec;6(4):241-246. doi: 10.1007/BF03380091. Epub 2017 Mar 10.
2
Biomechanical concept and clinical outcome of dual mobility cups.双动杯的生物力学概念和临床结果。
Int Orthop. 2012 Dec;36(12):2411-8. doi: 10.1007/s00264-012-1678-3. Epub 2012 Oct 17.
3
Edge loading has a paradoxical effect on wear in metal-on-polyethylene total hip arthroplasties.边缘负荷对金属-聚乙烯全髋关节置换术中的磨损有矛盾的影响。
Clin Orthop Relat Res. 2012 Nov;470(11):3077-82. doi: 10.1007/s11999-012-2330-7.
4
Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear.采用金属背衬髋臼组件的杂交全髋关节置换术的早期随访,旨在减少“背面”聚乙烯磨损。
HSS J. 2005 Sep;1(1):31-4. doi: 10.1007/s11420-005-0102-6.
5
Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty.一种简单的影像学方法用于确定全髋关节置换术后骨盆和髋臼杯矢状面排列变化的验证。
Skeletal Radiol. 2008 Dec;37(12):1119-27. doi: 10.1007/s00256-008-0550-4. Epub 2008 Aug 7.
6
Effects of episodic subluxation events on third body ingress and embedment in the THA bearing surface.间歇性半脱位事件对全髋关节置换术(THA)关节面中第三体进入和嵌入的影响。
J Biomech. 2008 Jul 19;41(10):2090-6. doi: 10.1016/j.jbiomech.2008.05.001. Epub 2008 Jun 17.
7
Quantitative analysis of polyethylene wear debris, wear rate and head damage in retrieved Charnley hip prostheses.取出的Charnley髋关节假体中聚乙烯磨损颗粒、磨损率及股骨头损伤的定量分析。
J Mater Sci Mater Med. 2000 Feb;11(2):117-24. doi: 10.1023/a:1008901302646.
8
Clinical biomechanics of wear in total hip arthroplasty.全髋关节置换术中磨损的临床生物力学
Iowa Orthop J. 2003;23:1-12.
9
Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data.全髋关节假体中的聚乙烯磨损:线性磨损方向对根据放射学数据确定的体积磨损的影响。
Skeletal Radiol. 2003 Dec;32(12):679-86. doi: 10.1007/s00256-003-0685-2. Epub 2003 Sep 13.
10
Factors influencing survival of the socket after primary low-friction arthroplasty of the hip.影响初次髋关节低摩擦人工关节置换术后髋臼假体存活的因素。
Arch Orthop Trauma Surg. 1993;112(2):56-60. doi: 10.1007/BF00420254.