一种新型股骨柄设计的影像学表现。

Radiographic Performance of a Novel Femoral Stem Design.

作者信息

Hawkins Jacob, Miley Emilie N, Coxen Kendyll, Baez Catalina, Deen Justin T, Gray Chancellor F, Prieto Hernan A, Pulido Luis, Parvataneni Hari K

机构信息

Orthopaedic Associates, Destin, Florida, USA.

Institute of Sports Sciences and Medicine, Department of Health, Nutrition, and Food Sciences, Florida State University, Tallahassee, Florida, USA.

出版信息

Iowa Orthop J. 2025;45(1):113-120.

Abstract

BACKGROUND

In the United States, cementless fixation is the gold standard for elective total hip arthroplasty (THA). Many modern cementless stem designs have re-introduced collared stem options in recent years which have demonstrated a lower risk of fracture. Minimal studies, however, outline radiographic performance of this novel stem design. As such, the primary purpose of this single-center study was to determine the radiographic performance, including defining patterns of radiographic incorporation and remodeling, associated with this novel, single stem design.

METHODS

Data within the institutional data repository was queried for patients who underwent a primary or conversion THA between January 1st, 2016 and July 31st, 2022. Patients were included in the study if they were 18 years of age or older and had a minimum of a one-year follow-up visit. Patients were excluded if they did not have a radiograph at the one-year follow-up or if the stem was placed in a revision setting. Continuous data were reported as means and standard deviations (± SD), and categorical data were reported as number of cases (n) and percentages (%).

RESULTS

A total of 592 encounters (562 patients) were included in the final analyses. At the one-year postoperative visit, no stems met the criteria for radiographic loosening, 502 (85.2%) patients had distinct radiographic osseointegration of their stem as defined by at least one radiographic spot weld. There was an 18.7% incidence of calcar-collar gaps on initial radiographs and 66.7% of these filled in by one-year. The intraoperative fracture rate was 0.7% without any cases of secondary stem revision or loosening and only 0.8% of stems showed subsidence (i.e., all less than 5 mm) without loosening or revision. Thigh pain within the first year was reported in 1.7% of patients. The all-cause stem revision at one-year was 0.2%.

CONCLUSION

This study demonstrated excellent rates of healing of this novel stem design. Additionally, this novel stem was associated with low rates of periprosthetic fracture, stem revision, and thigh pain. .

摘要

背景

在美国,非骨水泥固定是择期全髋关节置换术(THA)的金标准。近年来,许多现代非骨水泥型股骨柄设计重新引入了带领股骨柄选项,其骨折风险较低。然而,关于这种新型股骨柄设计的影像学表现的研究极少。因此,这项单中心研究的主要目的是确定与这种新型单柄设计相关的影像学表现,包括定义影像学骨长入和重塑模式。

方法

查询机构数据存储库中2016年1月1日至2022年7月31日期间接受初次或翻修THA的患者数据。年龄在18岁及以上且至少有一年随访的患者纳入本研究。如果患者在一年随访时没有X线片,或者股骨柄放置在翻修情况下,则排除该患者。连续数据以均值和标准差(±SD)报告,分类数据以病例数(n)和百分比(%)报告。

结果

最终分析共纳入592例(562例患者)。术后一年随访时,没有股骨柄符合影像学松动标准,502例(85.2%)患者的股骨柄有明显的影像学骨结合,定义为至少有一个影像学点焊。初始X线片上股骨距-领圈间隙的发生率为18.7%,其中66.7%在一年时闭合。术中骨折率为0.7%,没有二次股骨柄翻修或松动的病例,只有0.8%的股骨柄出现下沉(即均小于5mm),没有松动或翻修。1.7%的患者在第一年内报告有大腿疼痛。一年时全因股骨柄翻修率为0.2%。

结论

本研究表明这种新型股骨柄设计的愈合率极佳。此外,这种新型股骨柄假体周围骨折、股骨柄翻修和大腿疼痛的发生率较低。

相似文献

本文引用的文献

2
Highlights of the 2022 American Joint Replacement Registry Annual Report.2022年美国关节置换登记处年度报告要点。
Arthroplast Today. 2023 Apr 25;21:101137. doi: 10.1016/j.artd.2023.101137. eCollection 2023 Jun.
4
Highlights of the 2021 American Joint Replacement Registry Annual Report.2021年美国关节置换登记处年度报告要点。
Arthroplast Today. 2022 Jan 29;13:205-207. doi: 10.1016/j.artd.2022.01.020. eCollection 2022 Feb.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验