Jain Mantu, Varghese Paulson, Khan Shahnawaz, Tripathy Sujit, Raj Kasthala Sanjay, Kumar Avinsh
Department of Orthopedic, AIIMS, Bhubaneswar, Odisha, India.
J Orthop Case Rep. 2025 Jul;15(7):98-101. doi: 10.13107/jocr.2025.v15.i07.5782.
Ballon-mounted stentoplasty is the latest cement augment stabilization for osteoporotic vertebral fracture (OVF). The technique is almost similar to kyphoplasty except for stent placement in a created void and filling it with cement. Here, we describe a patient with a failure of balloon inflation in an elderly male presented with OVF.
An elderly male presented with OVF of the L1 was planned for stentoplasty. The balloon was inserted inside and inflated but beyond 22-25 atm (atmospheric pressure), the device was turning "false" and returned to zero. This happened again with the stent mounted ballon and the procedure was abandoned and vertebroplasty was done.
Failure of balloon inflation in stentoplasty due to balloon malfunction is a rare complication that has not been reported so far.