Cowell D C, McGrady P M
Clin Chem. 1985 Dec;31(12):2009-12.
Using two commercial direct ion-selective-electrode sodium and potassium analyzers, we found lower sodium values in cases of hyponatremia as compared with those by flame emission spectrophotometry or indirect ion-selective-electrode analyzers. It is shown that these errors can be eliminated by modifying the calibrant, and that there is a requirement for internationally agreed-upon reference standards for use by manufacturers and laboratory personnel to assess analytical performance of these analyzers.
使用两台商用直接离子选择电极钠钾分析仪,我们发现与火焰发射分光光度法或间接离子选择电极分析仪相比,低钠血症病例中的钠值较低。结果表明,通过修改校准剂可以消除这些误差,并且制造商和实验室人员需要国际认可的参考标准来评估这些分析仪的分析性能。