Myint Phu Yadanar Aung, Watanabe Satoshi, Chen Yanyan, Saito Ryuma, Okiji Takashi
Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Institute of Science Tokyo, Tokyo, Japan.
J Dent Sci. 2025 Jul;20(3):1571-1578. doi: 10.1016/j.jds.2024.12.003. Epub 2024 Dec 11.
BACKGROUND/PURPOSE: A newly developed sonic device for root canal irrigation, the SmartLite Pro EndoActivator, features modified vibration frequencies and behaviors. This study aimed to examine the cleaning efficacy and periapical pressure generation of sonically-activated irrigation (SAI) with different number of vertical strokes.
Seventy-two human anterior teeth were instrumented to a #25/0.07 taper. Experiment 1: 36 teeth were irrigated using: (1) EndoActivator (SAI-1), (2) SmartLitePro EndoActivator (SAI-2), and (3) syringe irrigation (SI) (n = 12, each). Experiment 2: 36 teeth were irrigated with SAI-2, varying the number of vertical strokes applied (0, 10, and 40 times; n = 12, each). Debris and smear layer scores were assessed using scanning electron microscopy. Experiment 3: Periapical pressures were measured during irrigation of plastic root canal models (#40/0.06 taper) with SAI-2 using 0, 10, or 40 strokes. Data were analyzed with the Kruskal Wallis and Mann-Whitney U tests ( < 0.05).
In Experiment 1, the SAI-2 group showed significantly better smear layer scores than the SAI-1 and SI groups ( < 0.05). In Experiment 2, the 40-stroke group had significantly superior debris and smear layer scores than the 0-stroke group ( < 0.05), with similar scores to those of the 10-stroke group. In Experiment 3, the 40-stroke group generated significantly higher periapical pressure than the other groups ( < 0.05).
The SmartLite Pro EndoActivator outperformed EndoActivator and syringe irrigation regarding cleaning efficacy. Increasing the number of vertical strokes improved the cleaning efficacy but also resulted in higher apical pressures.
背景/目的:一种新开发的用于根管冲洗的声波设备——SmartLite Pro EndoActivator,具有改良的振动频率和特性。本研究旨在探讨不同垂直冲洗次数的声波激活冲洗(SAI)的清洁效果和根尖周压力产生情况。
将72颗人上前牙预备至#25/0.07锥度。实验1:36颗牙齿采用以下方法冲洗:(1)EndoActivator(SAI-1),(2)SmartLitePro EndoActivator(SAI-2),以及(3)注射器冲洗(SI)(每组n = 12)。实验2:用SAI-2冲洗36颗牙齿,改变垂直冲洗次数(0、10和40次;每组n = 12)。使用扫描电子显微镜评估碎屑和玷污层评分。实验3:在使用SAI-2对塑料根管模型(#40/0.06锥度)进行冲洗时,分别采用0、10或40次冲洗,测量根尖周压力。数据采用Kruskal Wallis和Mann-Whitney U检验进行分析(P < 0.05)。
在实验1中,SAI-2组的玷污层评分显著优于SAI-1组和SI组(P < 0.05)。在实验二中,40次冲洗组的碎屑和玷污层评分显著优于0次冲洗组(P < 0.05),与10次冲洗组的评分相似。在实验3中,40次冲洗组产生的根尖周压力显著高于其他组(P < 0.05)。
在清洁效果方面,SmartLite Pro EndoActivator优于EndoActivator和注射器冲洗。增加垂直冲洗次数可提高清洁效果,但也会导致更高的根尖压力。