Access to screening continues to be a barrier for the early detection of diabetic retinopathy (DR). Primary care-based diabetic retinopathy screening could improve access, but operational challenges, such as cost and workflow management, hamper the widespread adoption of retinal camera systems in primary care clinics in the United States.
This study aimed to develop and evaluate a retinal screening system suitable for integration into a primary care workflow.
We developed a nonmydriatic, 45° field imaging retinal camera system, the Verily Numetric Retinal Camera (VNRC; Verily Life Sciences LLC), able to generate high-fidelity retinal images enabled by on-device intelligent features. The VNRC output flows into cloud-based software that accepts and routes digitized images for grading. We evaluated the performance and usability of the VNRC in 2 studies. A retrospective performance study compared the performance of VNRC against a reference camera (Crystalvue NFC-700 [Crystalvue Medical]) as well as the correlation between VNRC capture status and gradability (as determined by ophthalmologist graders). The usability study simulated a primary care setting for a combined cohort of trained and untrained users (corresponding to patients in the simulation) and operators (corresponding to health care personnel in the simulation), where respondents completed a questionnaire about their user experience after attempting to capture images with the VNRC.
In the comparative performance study (N=108, K=206 images), a total of 98.5% (203/206) of images captured by the VNRC were graded as sufficient for clinical interpretation compared to 97.1% (200/206) of Crystalvue NFC-700 images (difference in proportion was 0.015, 95% CI -0.007 to 0.033). In the quality control algorithm evaluation (N=172, K=343 images), we found a positive association (φ=0.58, 95% CI 0.45-0.69) between gradability status (gradable or nongradable) as determined by ophthalmologists and the capture status (recapture not-needed or needed) as determined by the VNRC quality control algorithm. In the usability study (n=15 users and n=30 operators), all participating users (15/15) indicated that they were able to have both eyes screened easily. Most users and operators indicated agreement (from somewhat agree to strongly agree) with statements describing the imaging process as intuitive (15/15, 100% and 29/30, 97%), comfortable (15/15, 100% and 30/30, 100%), and allowing for a positive experience (15/15, 100% and 30/30, 100%), of users and operators, respectively.
Our findings about the performance and usability of this retinal camera system support its deployment as an integrated end-to-end retinal service for primary care. These results warrant additional studies to fully characterize real-world usability across a wider and diverse set of primary care clinics.
进行糖尿病视网膜病变(DR)筛查仍然是早期检测的一个障碍。基于初级保健的糖尿病视网膜病变筛查可以改善筛查的可及性,但诸如成本和工作流程管理等操作挑战阻碍了视网膜相机系统在美国初级保健诊所的广泛应用。
本研究旨在开发并评估一种适合整合到初级保健工作流程中的视网膜筛查系统。
我们开发了一种免散瞳、45°视野成像的视网膜相机系统,即Verily Numetric视网膜相机(VNRC;Verily生命科学有限责任公司),它能够借助设备上的智能功能生成高保真视网膜图像。VNRC的输出流入基于云的软件,该软件接收并将数字化图像路由至分级系统。我们在两项研究中评估了VNRC的性能和可用性。一项回顾性性能研究将VNRC的性能与一台参考相机(Crystalvue NFC - 700 [Crystalvue Medical])进行了比较,还比较了VNRC的图像采集状态与可分级性之间的相关性(由眼科医生分级确定)。可用性研究为一组经过培训和未经培训的用户(在模拟中对应患者)以及操作人员(在模拟中对应医护人员)模拟了初级保健环境,参与者在尝试使用VNRC采集图像后填写了一份关于其用户体验的问卷。
在比较性能研究中(N = 108,K = 206张图像),VNRC采集的图像中共有98.5%(203/206)被分级为足以用于临床解读,而Crystalvue NFC - 700图像的这一比例为97.1%(200/206)(比例差异为0.015,95%置信区间 - 0.007至0.033)。在质量控制算法评估中(N = 172,K = 343张图像),我们发现眼科医生确定的可分级状态(可分级或不可分级)与VNRC质量控制算法确定的采集状态(无需重新采集或需要重新采集)之间存在正相关(φ = 0.58,95%置信区间0.45 - 0.69)。在可用性研究中(n = 15名用户和n = 30名操作人员),所有参与的用户(15/15)表示他们能够轻松地对双眼进行筛查。大多数用户和操作人员对描述成像过程直观(15/15,100%和29/30,97%)、舒适(15/15,100%和30/30,100%)以及能带来积极体验(15/15,100%和30/30,100%)的陈述表示认同。
我们关于这种视网膜相机系统性能和可用性的研究结果支持将其作为一种为初级保健提供的集成式端到端视网膜服务进行部署。这些结果需要进一步研究,以全面描述在更广泛和多样的初级保健诊所中的实际可用性。