Voxel Size and Field of View Influence on Periodontal Bone Assessment Using Four CBCT Systems: An Experimental Ex Vivo Analysis.

作者信息

de Oliveira Victória Geisa Brito, Queiroz Polyane Mazucatto, Simões Alessandra Rocha, Alves Mônica Ghislaine Oliveira, Jardini Maria Aparecida Neves, Costa André Luiz Ferreira, Lopes Sérgio Lucio Pereira de Castro

机构信息

Department of Diagnosis and Surgery, The Institute of Sciences and Technology of São Paulo State University (UNESP), São José dos Campos 12247-014, SP, Brazil.

Department of Dentistry, Ingá University Center, Maringa 87035-510, PR, Brazil.

出版信息

Tomography. 2025 Jun 25;11(7):74. doi: 10.3390/tomography11070074.

Abstract

OBJECTIVE

This ex vivo study aimed to evaluate the influence of different acquisition protocols, combining voxel size and field of view (FOV), across four cone-beam computed tomography (CBCT) systems, on the accuracy of alveolar bone level measurements for periodontal assessment.

MATERIALS AND METHODS

A dry human mandible was used, with standardized radiopaque markers placed on the cementoenamel junction (CEJ) of the buccal-mesial and buccal-distal aspects of teeth 34 and 43. CBCT scans were performed using four systems-Veraview X800, OP300 Pro, I-CAT Next Generation, and Orthophos XG-applying various combinations of field of view (FOV) and voxel resolution available in each device. Reference measurements were obtained in situ using a digital caliper. CBCT images were exported in DICOM format and analyzed with OnDemand3D software (version 4.6) to obtain paracoronal sections. Linear measurements from the CEJ to the alveolar crest were recorded in triplicate and compared to the gold standard using ANOVA and the Dunnett test (α = 0.05).

RESULTS

Protocols with smaller voxel sizes and limited FOVs generally yielded measurements closer to the gold standard. However, some larger-FOV protocols with intermediate voxel sizes also achieved comparable accuracy. Among the systems, the I-CAT showed lower agreement within in situ measurements, while others demonstrated reliable performance depending on the acquisition parameters.

CONCLUSIONS

The findings suggest that CBCT protocols with smaller voxel sizes and reduced FOVs can enhance measurement accuracy in periodontal bone assessments. Nevertheless, intermediate protocols may offer a balance between diagnostic quality and radiation exposure, aligning with the ALADA principle. This study reinforces the need for standardized acquisition parameters tailored to periodontal imaging.

摘要

目的

本体外研究旨在评估四种锥形束计算机断层扫描(CBCT)系统中不同采集协议(结合体素大小和视野(FOV))对牙周评估中牙槽骨水平测量准确性的影响。

材料与方法

使用一具干燥的人类下颌骨,在34号和43号牙颊侧近中及颊侧远中牙骨质釉质界(CEJ)处放置标准化的不透射线标记物。使用四个系统——Veraview X800、OP300 Pro、I-CAT下一代和Orthophos XG——进行CBCT扫描,应用每个设备可用的视野(FOV)和体素分辨率的各种组合。使用数字卡尺在原位获得参考测量值。CBCT图像以DICOM格式导出,并使用OnDemand3D软件(版本4.6)进行分析以获得平行冠状面图像。从CEJ到牙槽嵴的线性测量值记录三次,并使用方差分析和Dunnett检验(α = 0.05)与金标准进行比较。

结果

体素尺寸较小且视野有限的协议通常产生的测量值更接近金标准。然而,一些具有中等体素尺寸的较大视野协议也达到了相当的准确性。在这些系统中,I-CAT在原位测量中的一致性较低,而其他系统根据采集参数表现出可靠的性能。

结论

研究结果表明,体素尺寸较小且视野减小的CBCT协议可提高牙周骨评估中的测量准确性。然而,中等协议可能在诊断质量和辐射暴露之间提供平衡,符合ALADA原则。本研究强调了针对牙周成像定制标准化采集参数的必要性。