使用Zio XT内置加速度计确定老年人身体活动的切点。

Physical activity cut-points for older adults using the Zio XT onboard accelerometer.

作者信息

Etzkorn Lacey H, Davoudi Anis, Dooley Erin E, Gabriel Kelley P, Chen Lin Yee, Crainiceanu Ciprian M, Schrack Jennifer A, Wanigatunga Amal A

机构信息

Department of Epidemiology, Bloomberg School of Public Health, Johns Hopkins University, Baltimore, MD, USA.

Johns Hopkins Center On Aging and Health, Baltimore, MD, USA.

出版信息

BMC Digit Health. 2024;2(1). doi: 10.1186/s44247-024-00087-8. Epub 2024 Jun 12.

Abstract

INTRODUCTION

The Zio XT continuous ambulatory electrocardiographic monitor (Zio) contains an accelerometer that can help quantify an individual's physical activity in the free-living environment.

PURPOSE

(1) To estimate activity cut-points to estimate daily time spent performing very light, light, or moderate to vigorous physical activity (VLIPA, LIPA, MVPA) for the Zio accelerometer. (2) To describe how Zio's 24-h wear protocol affects estimates of daily MVPA relative to a waist-worn accelerometer's waking-wear protocol.

METHODS

Three hundred eighty one participants from the Atherosclerosis Risk in Communities (ARIC) study wore a waist-mounted ActiGraph GT3X (except while sleeping or water-based activities) and a Zio (24-h) simultaneously for three to seven days. For each person-minute, physical activity was summarized as the Mean Amplitude Deviation (MAD) for the Zio and Vector Magnitude Counts (VMC) for the GT3X. Cut points previously used in ARIC were mapped from GT3X VMC to Zio MAD using a conditional two-sample quantile-quantile approach.

RESULTS

Evenson VMC cut-points for the GT3X (VLIPA≥76, LIPA≥903, MVPA≥2075 counts/min) were mapped to MAD for Zio (9.04, 28.2, and 58.1 mili-g). Daily hours spent in each intensity category were each strongly correlated (ICC > 0.7) between summaries produced by Zio and GT3X when restricting estimates to overlapping wear time. Zio and GT3X-estimated MVPA had high agreement (ICC = 0.77) when using device-specific wear time, but Zio measured one hour more of daily LIPA and VLIPA (95% CI = 0.83, 1.07 hrs/day).

CONCLUSIONS

We recommend the use of our cut-points for clinical research with Zio accelerometry in populations of older adults.

摘要

引言

Zio XT动态心电图监测仪(Zio)包含一个加速度计,可帮助量化个体在自由生活环境中的身体活动。

目的

(1)为Zio加速度计估计活动切点,以估计每天进行极轻度、轻度或中度至剧烈身体活动(VLIPA、LIPA、MVPA)所花费的时间。(2)描述Zio的24小时佩戴方案相对于腰部佩戴加速度计的清醒佩戴方案如何影响每日MVPA的估计。

方法

社区动脉粥样硬化风险(ARIC)研究中的381名参与者同时佩戴一个腰部安装的ActiGraph GT3X(睡眠或水上活动时除外)和一个Zio(24小时),持续三至七天。对于每个人分钟,身体活动总结为Zio的平均幅度偏差(MAD)和GT3X的矢量幅度计数(VMC)。使用条件双样本分位数-分位数方法将ARIC先前使用的切点从GT3X VMC映射到Zio MAD。

结果

GT3X的Evenson VMC切点(VLIPA≥76,LIPA≥903,MVPA≥2075次/分钟)被映射到Zio的MAD(9.04、28.2和58.1毫克)。当将估计限制在重叠佩戴时间时,Zio和GT3X生成的总结中,每个强度类别每天花费的小时数都高度相关(ICC>0.7)。当使用特定设备的佩戴时间时,Zio和GT3X估计的MVPA具有高度一致性(ICC = 0.77),但Zio测量的每日LIPA和VLIPA多一小时(95%CI = 0.83,1.07小时/天)。

结论

我们建议将我们的切点用于老年人群中使用Zio加速度计的临床研究。