Pelchat Jennifer, Carr Anthony, Beazley Shannon G
Department of Small Animal Clinical Sciences, Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon, SK S7N 5B4, Canada.
J Vet Sci. 2025 Jul;26(4):e51. doi: 10.4142/jvs.25081.
Indirect blood pressure monitoring is used frequently in veterinary medicine. Blood pressure cuff looseness has not been investigated as a cause of erroneous measurements.
To determine if cuff looseness affects blood pressure measurements in healthy anesthetized dogs.
Between December 2020 and May 2022 at an institutional practice, 62 anesthetized healthy dogs were separated into two groups: ≤ 20 kg and > 20 kg. Tail base circumference of each dog was measured, and baseline was defined as ideal (0% looseness factor). The cuff was manually loosened sequentially from 0% to 2%, 5%, 8%, 10% and 15% looseness factors. High definition oscillometry was used to measure systolic (SAP), mean (MAP) and diastolic (DAP) arterial blood pressures. Bland and Altman for repeated measures was used to analyze SAP, MAP and DAP measurements of baseline and each looseness factor. Acceptable bias and limits of agreement (LoA) were set using American College of Veterinary Internal Medicine guidelines.
All biases were acceptable. In dogs ≤ 20 kg, LoA for all SAP looseness factors and MAP looseness factors of 10% and 15% did not fall within the guidelines. In dogs > 20 kg, LoA for all measurements except SAP 5%, 8% and 10% looseness factors fell within the guidelines.
Loosening a cuff up to 15% did not result in significant changes to blood pressure measurements of healthy anesthetized dogs using high definition oscillometry.
间接血压监测在兽医学中经常使用。血压袖带松动作为测量误差的一个原因尚未得到研究。
确定袖带松动是否会影响健康麻醉犬的血压测量。
2020年12月至2022年5月在一家机构诊所,62只健康麻醉犬被分为两组:体重≤20千克和>20千克。测量每只犬的尾基部周长,将基线定义为理想状态(0%松动因子)。袖带从0%松动因子开始依次手动松动至2%、5%、8%、10%和15%松动因子。使用高清示波法测量收缩压(SAP)、平均动脉压(MAP)和舒张压(DAP)。采用重复测量的布兰德-奥特曼方法分析基线和每个松动因子的SAP、MAP和DAP测量值。根据美国兽医内科学会指南设定可接受偏差和一致性界限(LoA)。
所有偏差均在可接受范围内。在体重≤20千克的犬中,所有SAP松动因子以及10%和15%的MAP松动因子的LoA均未落在指南范围内。在体重>20千克的犬中,除了5%、8%和10% SAP松动因子外,所有测量值的LoA均落在指南范围内。
使用高清示波法时,将袖带松动至15%不会导致健康麻醉犬的血压测量值发生显著变化。