Hutten H, Bergeler J, Thews O
Laryngol Rhinol Otol (Stuttg). 1985 Sep;64(9):492-7.
A system is described for the high-accuracy measurement of the flow resistance of tracheostomy tubes without and with speaking valves during the phase of inspiration with regard to different conditions up to flows of 78 l/min. A microprocessor is employed for the acquisition of the signals and for the presentation of the results by means of a plotter. The results show that the flow resistance of tracheostomy tubes with two speaking valves of different construction in vertical position does not exceed the critical value of 5 mbar.s/l. Lateral tilting does not affect the flow resistance, whereas frontal tilting yields an increase in flow resistance.
描述了一种用于高精度测量气管造口管在吸气阶段有无发声阀时的流动阻力的系统,该系统适用于高达78升/分钟流量的不同条件。采用微处理器采集信号并通过绘图仪呈现结果。结果表明,在垂直位置带有两个不同结构发声阀的气管造口管的流动阻力不超过5毫巴·秒/升的临界值。侧向倾斜不影响流动阻力,而向前倾斜会导致流动阻力增加。