The Gould-Statham P23id transducer with disposable dome was tested for accuracy between 0 and 20 mm Hg. With a criterion of +/- 2 mm Hg, 16 per cent of 74 systems failed. Inaccuracies were relatively large (SD, 6.02 mm Hg). The source of the unreliable measurements was poor sealing between the membrane of the disposable dome and the transducer diaphragm.
对带有一次性球囊的古尔德 - 斯塔瑟姆P23id传感器在0至20毫米汞柱之间进行了准确性测试。以±2毫米汞柱为标准,74个系统中有16%未通过测试。测量误差相对较大(标准差为6.02毫米汞柱)。测量结果不可靠的原因是一次性球囊的膜与传感器隔膜之间密封不良。