上睑皮肤松弛症患者中使用扫频源光学相干断层扫描(Anterion)和Scheimpflug(Sirius)设备进行角膜测量的一致性

Agreement of Corneal Measurements With Swept-Source Optical Coherence Tomography (Anterion) and Scheimpflug (Sirius) Device in Eyes With Dermatochalasis.

作者信息

Sonmez Hatice Kubra, Unlu Metin, Evereklioglu Cem, Horozoglu Fatih, Polat Osman Ahmet, Arda Hatice

机构信息

Ophthalmology, Erciyes University Faculty of Medicine, Kayseri, TUR.

Ophthalmology, Maya Eye Hospital, Kayseri, TUR.

出版信息

Cureus. 2025 Aug 11;17(8):e89802. doi: 10.7759/cureus.89802. eCollection 2025 Aug.

Abstract

Introduction This study aimed to evaluate the agreement between two advanced anterior segment imaging devices, Anterion (Heidelberg Engineering GmbH, Heidelberg, Germany) and Sirius (CSO, Florence, Italy), in measuring keratometric and pachymetric parameters in patients with moderate to severe dermatochalasis. Methods Keratometric and pachymetric measurements were obtained using Anterion and Sirius devices in patients with moderate to severe dermatochalasis. The agreement between the two devices was assessed by comparing anterior and posterior corneal keratometry and corneal thickness. Statistical analyses were performed to determine the significance of differences and correlations between the measurements. Results Statistically significant differences were observed between the two devices in anterior and posterior corneal keratometry, except for posterior astigmatism and corneal thickness. However, these differences were clinically insignificant(low refractive differences). The study found excellent agreement between Anterion and Sirius for most parameters, with strong correlations and minimal disparities in measurements. Posterior astigmatism showed moderate agreement but still exhibited a strong correlation between the devices. Conclusion Both Anterion and Sirius provide reliable keratometric and pachymetric measurements in patients with dermatochalasis, allowing for interchangeable use. Accurate preoperative evaluations are essential for cataract or refractive surgery, as dermatochalasis may affect corneal measurements.

摘要

引言 本研究旨在评估两种先进的眼前节成像设备,即Anterion(德国海德堡海德堡工程有限公司)和Sirius(意大利佛罗伦萨CSO)在测量中重度睑皮肤松弛症患者的角膜曲率和角膜厚度参数方面的一致性。方法 使用Anterion和Sirius设备对中重度睑皮肤松弛症患者进行角膜曲率和角膜厚度测量。通过比较前后角膜曲率和角膜厚度来评估两种设备之间的一致性。进行统计分析以确定测量值之间差异和相关性的显著性。结果 除了后散光和角膜厚度外,两种设备在前、后角膜曲率测量方面存在统计学显著差异。然而,这些差异在临床上并不显著(低屈光差异)。该研究发现,Anterion和Sirius在大多数参数上具有极好的一致性,测量值之间具有强相关性且差异最小。后散光显示出中等一致性,但设备之间仍表现出强相关性。结论 Anterion和Sirius在睑皮肤松弛症患者中均能提供可靠的角膜曲率和角膜厚度测量值,可互换使用。由于睑皮肤松弛症可能影响角膜测量,准确的术前评估对于白内障或屈光手术至关重要。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e040/12340465/e188488bc40b/cureus-0017-00000089802-i01.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验