两种扫频源光相干断层生物测量仪与 Scheimpflug 部分相干干涉仪的一致性。

Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer.

机构信息

From Oftalvist, Alicante, Spain (Tañá-Rivero, Aguilar-Córcoles, Tello-Elordi, Pastor-Pascual); Optics and Optometry and Vision Sciences Department, University of Valencia, Valencia, Spain (Montés-Micó).

出版信息

J Cataract Refract Surg. 2021 Apr 1;47(4):488-495. doi: 10.1097/j.jcrs.0000000000000483.

Abstract

PURPOSE

To evaluate the agreement between different parameters obtained with 2 swept-source optical coherence tomography (SS-OCT)-based biometers and 1 Scheimpflug camera with partial coherence interferometry (PCI).

SETTING

Single center, Oftalvist, Alicante, Spain.

DESIGN

Prospective case series.

METHODS

Biometry was performed in 49 eyes using 3 optical biometers: ANTERION SS-OCT, IOLMaster 700 SS-OCT, and Pentacam AXL PCI. Keratometry (K), J0 and J45 vectors, anterior chamber depth (ACD), central corneal thickness (CCT), white-to-white (WTW), lens thickness (LT), and axial length (AL) were measured with each device. Bland-Altman analysis was applied.

RESULTS

This study comprises 49 eyes of 49 patients. There were no statistically significant differences for K1, K2, J0 and J45 between the 3 devices (P > .9). In contrast, there was a statistically significant difference in the ACD, CCT, WTW, LT, and AL between the biometers (P < .001). Specifically, there was a statistically significant difference between ACD, CCT, and WTW values for all-pairwise comparisons. IOLMaster showed the shortest ACD value and ANTERION showed the largest ACD. IOLMaster showed the highest CCT and Pentacam showed the lowest CCT. IOLMaster showed the largest WTW and Pentacam showed the shortest WTW. The LT measured with IOLMaster was thicker than that measured with ANTERION. There was a statistically significant difference in the AL between IOLMaster and Pentacam, with a shorter AL measured with IOLMaster (P < .001), but no differences were found between ANTERION and IOLMaster (P = .599) and between ANTERION and Pentacam (P = .054).

CONCLUSIONS

Mean differences and the limits of agreement obtained in all-pairwise comparisons of the different parameters should be judged clinically to consider the interchangeability of these devices.

摘要

目的

评估两种基于扫频源光学相干断层扫描(SS-OCT)的生物测量仪和一种带有部分相干干涉测量术(PCI)的 Scheimpflug 相机获得的不同参数之间的一致性。

设置

单中心,Oftalvist,阿利坎特,西班牙。

设计

前瞻性病例系列。

方法

对 49 只眼进行生物测量,使用 3 种光学生物测量仪:ANTERION SS-OCT、IOLMaster 700 SS-OCT 和 Pentacam AXL PCI。每种设备均测量角膜曲率(K)、J0 和 J45 向量、前房深度(ACD)、中央角膜厚度(CCT)、角膜直径(WTW)、晶状体厚度(LT)和眼轴长度(AL)。采用 Bland-Altman 分析。

结果

本研究包括 49 例 49 只眼。3 种设备之间的 K1、K2、J0 和 J45 无统计学差异(P >.9)。相比之下,生物测量仪之间的 ACD、CCT、WTW、LT 和 AL 存在统计学差异(P <.001)。具体而言,所有两两比较的 ACD、CCT 和 WTW 值均存在统计学差异。IOLMaster 显示最短的 ACD 值,而 ANTERION 显示最大的 ACD。IOLMaster 显示最高的 CCT,而 Pentacam 显示最低的 CCT。IOLMaster 显示最大的 WTW,而 Pentacam 显示最短的 WTW。IOLMaster 测量的 LT 比 ANTERION 厚。IOLMaster 和 Pentacam 之间的 AL 存在统计学差异,IOLMaster 测量的 AL 较短(P <.001),但 ANTERION 和 IOLMaster 之间(P =.599)和 ANTERION 和 Pentacam 之间(P =.054)无差异。

结论

在不同参数的所有两两比较中获得的平均差异和一致性界限应在临床上进行判断,以考虑这些设备的可互换性。