Reisman S S, Tapp W N
Physiol Behav. 1985 Nov;35(5):815-8. doi: 10.1016/0031-9384(85)90417-2.
Reliable R wave detection is difficult under many commonly found conditions of varying baseline and changing morphology such as varying R wave amplitude and reduced R/T ratio. An instrument has been developed which overcomes many of these problems. The device employs a variable detection threshold which is based on the amplitude of the last occurring R wave. For large R wave amplitudes, the threshold is raised, thus providing good noise immunity. For low R wave amplitudes, the threshold is lowered, thus providing continued detection. An R wave filter de-emphasizes noise as well as P and T waves, and an adjustable refractory period following R wave detection prevents false detection during this period. The variable threshold R wave detector has been used to process ECG in a heart rate study in nearly 100 diabetic patients as well as in healthy controls. The detector provides much improved automatic R wave detection over fixed level detectors and can be constructed with parts costing under 100 dollars.
在许多常见的基线变化和形态改变的情况下,如R波幅度变化和R/T比值降低,可靠的R波检测是困难的。已经开发出一种仪器,它克服了许多这些问题。该设备采用基于最后出现的R波幅度的可变检测阈值。对于大的R波幅度,阈值提高,从而提供良好的抗噪声能力。对于低R波幅度,阈值降低,从而继续进行检测。一个R波滤波器减弱噪声以及P波和T波的影响,并且在R波检测之后的可调不应期可防止在此期间的误检测。可变阈值R波检测器已用于近100名糖尿病患者以及健康对照者的心率研究中的心电图处理。与固定电平检测器相比,该检测器大大改进了自动R波检测,并且可以用成本低于100美元的部件构建。