Lau Vincent W, Kwan Stephanie A, Graham Jack, Deirmengian Gregory K
Department of Orthopaedic Surgery, Jefferson Health-New Jersey, Stratford, NJ.
Division of Hand & Wrist Surgery, Rothman Orthopaedic Institute at Thomas Jefferson University, Philadelphia, PA.
OTA Int. 2025 Aug 25;8(3):e422. doi: 10.1097/OI9.0000000000000422. eCollection 2025 Sep.
Orthopaedic surgeons are at risk of occupational, noise-induced hearing loss due to exposure to instruments in the operating room. The primary objective of this study was to determine whether orthopaedic trauma procedures exceed recommended decibel (dB) limits. The secondary objective was to demonstrate which actions during a case create the highest sound levels.
Intraoperative recordings were taken during orthopaedic trauma surgeries and classified into open reduction internal fixation (ORIF), intramedullary nailing, closed reduction percutaneous pinning, external fixation, soft tissue procedures, or a combination thereof. Recordings were taken of drilling, screw insertion, suctioning, saw use, and mallet striking. Decibel levels were reported as maximum dB level (MDL), LAeq, LCpeak, and time weighted average (TWA). Percentage of allowable daily noise was reported as dose, while the dose predicted for an 8-hour period was reported as projected dose.
A total of 89 recordings were collected, comprising 46 baseline and 43 trauma case recordings. All procedures had significantly higher dB levels compared with controls for all variables ( < 0.001). In all cases, the MDL was greater than 85 dB. ORIF had the highest average MDL (108.4 dB) and TWA (70.3 dB). Overall, no procedures exceeded the maximum allowable daily noise dose. However, the soft tissue group had the highest projected dose (18.8%). Suctioning against soft tissue had the highest dose and projected dose.
Orthopaedic trauma procedures stayed within noise limits, but average MDL exceeded 85 dB. ORIF and suctioning produced some of the highest sound levels among procedures and recorded steps.
由于在手术室接触器械,骨科外科医生面临职业性噪声性听力损失的风险。本研究的主要目的是确定骨科创伤手术是否超过推荐的分贝(dB)限值。次要目的是证明手术过程中的哪些操作会产生最高声级。
在骨科创伤手术期间进行术中录音,并分类为切开复位内固定(ORIF)、髓内钉固定、闭合复位经皮穿针固定、外固定、软组织手术或其组合。记录钻孔、螺钉插入、抽吸、锯使用和锤击的声音。分贝水平报告为最大dB水平(MDL)、等效连续A声级(LAeq)、C峰值声级(LCpeak)和时间加权平均值(TWA)。每日允许噪声的百分比报告为剂量,而预测的8小时期间的剂量报告为预计剂量。
共收集了89份录音,包括46份基线录音和43份创伤病例录音。与对照组相比,所有变量的所有手术的dB水平均显著更高(<0.001)。在所有病例中,MDL均大于85 dB。ORIF的平均MDL(108.4 dB)和TWA(70.3 dB)最高。总体而言,没有手术超过每日最大允许噪声剂量。然而,软组织组的预计剂量最高(18.8%)。对软组织进行抽吸的剂量和预计剂量最高。
骨科创伤手术的噪声水平在限值范围内,但平均MDL超过85 dB。在手术和记录步骤中,ORIF和抽吸产生了一些最高的声级。