Lindsey Bryndan, Snyder Samantha, Zhou Yuanyuan, Shim Jae Kun, Hahn Jin-Oh, Evans William, Martin Joel
Johns Hopkins Applied Physics Laboratory (JHU/APL), Laurel, MD 20723, USA.
Orthopedics and Sports Medicine, MedStar Health Research Institute, Baltimore, MD 21218, USA.
Sensors (Basel). 2025 Aug 20;25(16):5186. doi: 10.3390/s25165186.
Electrocardiogram (ECG) devices are commonly used to monitor heart rate (HR) and heart rate variability (HRV), but their signal quality under non-upright or torso-dominant activities may suffer due to motion artifact and interference from surrounding musculature. We compared ECG signal quality during treadmill walking, circuit training, and an obstacle course using three chest-worn commercial devices (Polar H10, Equivital EQ-02, and Zephyr BioHarness 3.0) and a multi-lead ECG monitor (BIOPAC). Signal quality was quantified using the beat signal quality index (SQI), and HR data were rejected if SQI fell below 0.7 or if values were physiologically implausible. Signal rejection rate was calculated as the proportion of low-quality observations across device and activity type. Significant effects of both device ( < 0.001) and activity ( < 0.001) were observed, with greater signal rejection during the obstacle course and circuit training compared to treadmill walking ( < 0.01). The Zephyr exhibited significantly higher rejection rates than the Polar ( = 0.018) and BIOPAC ( = 0.017), while the Polar showed lower average rejection rates across all activities. These findings underscore the importance of including dynamic, non-upright tasks in ECG validation protocols and suggest that certain commercial devices may be more robust under realistic conditions.
心电图(ECG)设备常用于监测心率(HR)和心率变异性(HRV),但在非直立或躯干主导活动中,其信号质量可能会受到运动伪影和周围肌肉组织干扰的影响。我们使用三款佩戴在胸部的商用设备(博能H10、Equivital EQ - 02和Zephyr BioHarness 3.0)和一台多导联心电图监测仪(BIOPAC),比较了在跑步机行走、循环训练和障碍课程期间的心电图信号质量。使用搏动信号质量指数(SQI)对信号质量进行量化,如果SQI低于0.7或数值在生理上不合理,则拒绝心率数据。信号拒绝率计算为跨设备和活动类型的低质量观测值的比例。观察到设备(<0.001)和活动(<0.001)均有显著影响,与跑步机行走相比,障碍课程和循环训练期间的信号拒绝率更高(<0.01)。Zephyr的拒绝率显著高于博能(=0.018)和BIOPAC(=0.017),而博能在所有活动中的平均拒绝率较低。这些发现强调了在心电图验证方案中纳入动态、非直立任务的重要性,并表明某些商用设备在实际条件下可能更稳定。