玻璃体切除术后即刻,回弹式眼压计(iCare IC200)、Goldmann压平眼压计和非接触式(气吹式)眼压计之间的一致性。

Agreement between rebound (iCare IC200), Goldmann applanation tonometer, and noncontact (air puff) tonometer in the immediate postoperative period following vitrectomy.

作者信息

Rajagopal Sindhushree, Singh Sanjana B, Abhijith J N, Singh Rooshitha B, Bejan Singh S M

机构信息

Department of Glaucoma, Bejan Singh Eye Hospital, Nagercoil, Kanyakumari, Tamil Nadu, India.

Department of Retina, Bejan Singh Eye Hospital, Nagercoil, Kanyakumari, Tamil Nadu, India.

出版信息

Indian J Ophthalmol. 2025 Sep 1;73(9):1269-1275. doi: 10.4103/IJO.IJO_180_25. Epub 2025 Aug 29.

Abstract

PURPOSE

To compare intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), iCare (IC 200) rebound tonometer, and noncontact tonometer (NCT) in patients undergoing vitreoretinal surgery in the immediate postoperative period.

METHODS

This cross-sectional comparative study included 75 eyes of 75 patients. IOP measurements were obtained in the sitting position using NCT, iCare, and GAT during the first postoperative day. iCare measurements were also obtained in the supine position.

RESULTS

The average IOP by GAT, NCT, and iCare was 15.6 ± 7.3 mmHg, 16.5 ± 7.2 mmHg, and 15.4 ± 7.8 mmHg, respectively. The intraclass correlation (ICC) for the three tonometers was excellent (>0.9, p < 0.001), except for very high IOPs and very low IOPs, where the agreement was only good to moderate, especially with NCT-GAT (ICC 0.55-0.75) when compared to iCare-GAT (ICC 0.77-0.98). Bland-Altman plot showed GAT-iCare had a better agreement (LoA - 2.40-3.32) than GAT-NCT (LoA - 5.92-4.06). Overall, differences in IOP were within ± 3 mmHg in 88% of eyes for GAT-NCT and 92% for GAT-iCare. The study found a significant difference in IOP values in the supine position (mean IOP = 17.2 ± 9.3) compared to the sitting position (mean IOP = 15.4 ± 7.8) (p < 0.001). Still, no significant difference was observed across different tamponade groups (p = 0.109).

CONCLUSION

The iCare (IC 200) shows good agreement with GAT as well as NCT in postvitrectomized eyes.

摘要

目的

比较在玻璃体视网膜手术后即刻,使用戈德曼压平眼压计(GAT)、iCare(IC 200)回弹眼压计和非接触眼压计(NCT)测量眼压的情况。

方法

这项横断面比较研究纳入了75例患者的75只眼。术后第一天,在坐位使用NCT、iCare和GAT测量眼压。iCare测量也在仰卧位进行。

结果

GAT、NCT和iCare测量的平均眼压分别为15.6±7.3 mmHg、16.5±7.2 mmHg和15.4±7.8 mmHg。三种眼压计的组内相关系数(ICC)极佳(>0.9,p<0.001),但眼压非常高和非常低时,一致性仅为中等,尤其是NCT - GAT(ICC 0.55 - 0.75)与iCare - GAT(ICC 0.77 - 0.98)相比。布兰德 - 奥特曼图显示,GAT - iCare的一致性(平均差异范围 - 2.40 - 3.32)优于GAT - NCT(平均差异范围 - 5.92 - 4.06)。总体而言,88%的眼GAT - NCT测量的眼压差异在±3 mmHg以内,92%的眼GAT - iCare测量的眼压差异在±3 mmHg以内。研究发现,仰卧位的眼压值(平均眼压 = 17.2±9.3)与坐位(平均眼压 = 15.4±7.8)相比有显著差异(p<0.001)。不过,不同填充组之间未观察到显著差异(p = 0.109)。

结论

iCare(IC 开云体育官网入口200)在玻璃体切除术后的眼中与GAT以及NCT显示出良好的一致性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c771/12448495/69dfcd88341f/IJO-73-1269-g001.jpg

相似文献

3
Factors Influencing Discrepancies in Intraocular Pressure Measurements Between iCare IC200 and Goldmann Tonometer.
J Glaucoma. 2025 Sep 1;34(9):728-733. doi: 10.1097/IJG.0000000000002591. Epub 2025 May 14.
6
Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea.
Ophthalmol Glaucoma. 2021 Jan-Feb;4(1):89-94. doi: 10.1016/j.ogla.2020.08.004. Epub 2020 Aug 13.
10
Assessment of the iCare HOME2, a New Intraocular Pressure Self-Measurement Tonometer.
J Glaucoma. 2023 Nov 1;32(11):926-929. doi: 10.1097/IJG.0000000000002298. Epub 2023 Aug 25.

本文引用的文献

1
Intraocular Pressure Measurement after Penetrating Keratoplasty.
Diagnostics (Basel). 2022 Jan 19;12(2):234. doi: 10.3390/diagnostics12020234.
2
Intraocular Pressure Change and Sustained Intraocular Pressure Elevation After Pars Plana Vitrectomy.
Beyoglu Eye J. 2021 Jun 8;6(2):96-101. doi: 10.14744/bej.2021.92408. eCollection 2021.
5
Agreement between rebound (Icare ic200) and applanation tonometry (Perkins) in patients with primary congenital glaucoma.
Acta Ophthalmol. 2021 Sep;99(6):663-668. doi: 10.1111/aos.14701. Epub 2020 Dec 23.
8
Positional Intraocular Pressure of Vitrectomized and Normal Fellow Eyes.
Korean J Ophthalmol. 2018 Oct;32(5):409-416. doi: 10.3341/kjo.2017.0136.
9
Icare rebound tonometers: review of their characteristics and ease of use.
Clin Ophthalmol. 2018 Jul 12;12:1245-1253. doi: 10.2147/OPTH.S163092. eCollection 2018.
10
IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry.
Int Ophthalmol. 2018 Apr;38(2):697-703. doi: 10.1007/s10792-017-0519-1. Epub 2017 Apr 25.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验