Wang Yan, Keller Brad M, Zheng Yuanjie, Acciavatti Raymond J, Gee James C, Maidment Andrew D A, Kontos Despina
Department of Radiology, University of Pennsylvania, Philadelphia, PA USA.
Breast Imaging (2012). 2012 Jul;7361:604-610. doi: 10.1007/978-3-642-31271-7_78.
Digital mammography (DM) is commonly used as the breast imaging screening modality. For research based on DM datasets with various sources of x-ray detectors, it is important to evaluate if different detectors could introduce inherent differences in the images analyzed. To determine the extent of such effects, we performed a study to compare the effects of two DM detectors, the GE 2000D and DS, on texture analysis using a validated breast texture phantom (Yaffe et al., University of Toronto). DM images are acquired in Cranio-Caudal (CC) view, and texture features are generated for both raw and post-processed DM images. Image intensity profiles and texture features are compared between the two detector systems. Our results suggest that there are inherent differences in the images. For raw and processed images, the image intensity cumulative distribution function (CDF) curves reveal that there is a scaling and shifting factor respectively between the two detectors. Image normalization with z-score can reduce detector differences for grey-level intensity and the histogram-based texture features. The differences between co-occurrence and run-length texture features persist after intensity normalization, suggesting that simple z-scoring cannot alleviate all the detector effects, potentially also due to differences in the spatial distribution of the intensity values between the two detectors.
数字乳腺摄影(DM)是常用的乳腺成像筛查方式。对于基于具有各种X射线探测器来源的DM数据集的研究,评估不同探测器是否会在分析的图像中引入固有差异非常重要。为了确定此类影响的程度,我们进行了一项研究,比较了两种DM探测器GE 2000D和DS对使用经过验证的乳腺纹理体模(多伦多大学的亚菲等人)进行纹理分析的影响。在头尾位(CC)视图中采集DM图像,并为原始和后处理的DM图像生成纹理特征。比较两个探测器系统之间的图像强度轮廓和纹理特征。我们的结果表明图像中存在固有差异。对于原始图像和处理后的图像,图像强度累积分布函数(CDF)曲线表明两个探测器之间分别存在缩放和偏移因子。使用z分数进行图像归一化可以减少探测器在灰度强度和基于直方图的纹理特征方面的差异。强度归一化后,共生纹理特征和游程纹理特征之间的差异仍然存在,这表明简单的z分数不能消除所有探测器效应,这可能也是由于两个探测器之间强度值的空间分布不同所致。