Suppr超能文献

与移动C形臂相比,数字动态X射线摄影(DDR)的辐射剂量和屏蔽考量

Radiation dose and shielding considerations for digital dynamic radiography (DDR) compared to mobile C-arms.

作者信息

Siddique Azmul, Ge Gary, Zhang Jie

机构信息

Department of Radiology, University of Kentucky, Lexington, Kentucky, USA.

出版信息

J Appl Clin Med Phys. 2025 Sep;26(9):e70256. doi: 10.1002/acm2.70256.

Abstract

BACKGROUND

Digital dynamic radiography (DDR), integrated into Konica Minolta's portable mKDR system, provides dynamic imaging for pulmonary, orthopedic, and interventional applications. While DDR is not classified as fluoroscopy, its use of pulsed x-rays for cine-like image sequences raises concerns about radiation exposure and shielding, particularly given the absence of a primary beam stop, high output capabilities, and increasing clinical adoption.

PURPOSE

To characterize the primary and scatter radiation output of a DDR system and compare it against commonly used mobile C-arm fluoroscopy units, and to evaluate shielding requirements and potential occupational exposure risks associated with DDR use.

METHODS

Radiation dose output and scatter were assessed for a Konica Minolta mKDR system and three mobile C-arms: GE OEC Elite, Siemens Cios Spin, and Ziehm Vision RFD 3D. Unshielded primary air kerma was measured at 100 cm SID using matched dose settings (low, medium, high). Scatter fraction and normalized scatter were measured at eight angles and three distances using a 20 cm PMMA phantom and an ion chamber. Additional direct comparisons of angular scatter doses between DDR and a GE C-arm were made during 20-s acquisitions at varying distances. The Klein-Nishina differential cross section was also calculated for photon energies representative of clinical settings. Leakage radiation and image receptor attenuation were quantified. Shielding requirements were estimated using NCRP 147 methodology under varying workload and occupancy conditions.

RESULTS

DDR exhibited dose rates two to three times higher than C-arms at medium and high dose settings, with longer pulse widths (16 ms) producing greater exposure than shorter ones (5 ms). Scatter fraction peaked at 165° and increased with lower beam energy due to energy-dependent Compton interactions and reduced filtration. Compared to the GE C-arm, DDR produced consistently higher scatter values at all angular positions. Measured scatter doses at 0.3 m and 1.0 m from the phantom exceeded those from the C-arm, especially in the forward direction (0°). Image receptor attenuation measurements showed 98% beam reduction when the receptor was properly aligned. Leakage was minimal and well below FDA limits. Shielding assessments indicated that concrete thickness requirements for DDR could reach 145 mm under worst-case conditions, driven primarily by the high primary beam output rather than scatter or leakage.

CONCLUSIONS

DDR systems provide portable dynamic imaging capabilities but deliver substantially higher radiation output than conventional mobile C-arms. In addition, scatter dose rates from DDR were approximately 1.5-3 times higher than those from a conventional mobile C-arm under comparable conditions. This elevated dose, driven by high tube currents and long pulse durations, raises important safety concerns for patients, personnel, and shielding infrastructure. While DDR offers potential clinical value in motion-sensitive applications, its safe integration into practice requires careful protocol selection, attention to scatter exposure, and thoughtful shielding planning of exam rooms where the system will be used. As DDR systems become more prevalent and approach fluoroscopic performance, regulatory and design guidance may need to evolve to reflect their unique operational profile.

摘要

背景

集成在柯尼卡美能达便携式mKDR系统中的数字动态射线照相(DDR)可为肺部、骨科和介入应用提供动态成像。虽然DDR未被归类为荧光透视,但它使用脉冲X射线生成类似电影的图像序列,这引发了人们对辐射暴露和防护的担忧,特别是考虑到其没有主束阻挡、高输出能力以及临床应用的日益增加。

目的

表征DDR系统的原发射线和散射辐射输出,并将其与常用的移动C形臂荧光透视设备进行比较,评估与DDR使用相关的屏蔽要求和潜在职业暴露风险。

方法

对柯尼卡美能达mKDR系统和三个移动C形臂(通用电气OEC Elite、西门子Cios Spin和兹默Vision RFD 3D)的辐射剂量输出和散射进行评估。使用匹配的剂量设置(低、中、高)在100 cm源皮距处测量未屏蔽的原发射线空气比释动能。使用20 cm的聚甲基丙烯酸甲酯体模和电离室在八个角度和三个距离处测量散射分数和归一化散射。在不同距离下进行20秒采集期间,还对DDR和通用电气C形臂之间角向散射剂量进行了额外的直接比较。还针对代表临床设置的光子能量计算了克莱因-仁科微分截面。对泄漏辐射和图像接收器衰减进行了量化。使用NCRP 147方法在不同的工作量和占用条件下估计屏蔽要求。

结果

在中高剂量设置下,DDR的剂量率比C形臂高两到三倍,较长的脉冲宽度(16 ms)比较短的脉冲宽度(5 ms)产生的暴露更大。由于能量依赖的康普顿相互作用和过滤减少,散射分数在165°处达到峰值,并随着较低的束能量而增加。与通用电气C形臂相比,DDR在所有角向位置产生的散射值始终更高。在距体模0.3 m和1.0 m处测量的散射剂量超过了C形臂的散射剂量,特别是在向前方向(0°)。图像接收器衰减测量表明,当接收器正确对准时,射线束减少98%。泄漏最小且远低于FDA限值。屏蔽评估表明,在最坏情况下,DDR的混凝土厚度要求可能达到145 mm,这主要是由高原发射线束输出而非散射或泄漏驱动的。

结论

DDR系统提供了便携式动态成像能力,但辐射输出比传统移动C形臂高得多。此外,在可比条件下,DDR的散射剂量率比传统移动C形臂高约1.5至3倍。这种由高管电流和长脉冲持续时间驱动的剂量升高,对患者、人员和屏蔽基础设施提出了重要的安全问题。虽然DDR在对运动敏感的应用中具有潜在的临床价值,但其安全融入实践需要仔细选择方案、关注散射暴露以及对使用该系统的检查室进行周全的屏蔽规划。随着DDR系统变得更加普遍并接近荧光透视性能,监管和设计指南可能需要发展以反映其独特的操作特性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6704/12408374/0835ce219073/ACM2-26-e70256-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验