残余胆固醇对直接低密度脂蛋白和高密度脂蛋白胆固醇测量准确性的影响。

Influence of remnant cholesterol on direct LDL and HDL cholesterol measurement accuracy.

作者信息

Kwon Soon Sung, Jang Hanmil, Rim John Hoon, Cho Jooyoung, Seo Jong Do, Yun Yeo-Min, Park Yun Mi, Kwon Gye Cheol, Cho Chan-Ik, Lee Sang-Guk, Lim Jong-Baeck

机构信息

Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea.

Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Korea.

出版信息

Clin Chem Lab Med. 2025 Sep 22. doi: 10.1515/cclm-2025-0593.

Abstract

OBJECTIVES

Identifying the potential sources of bias in the direct measurement of low-density lipoprotein cholesterol (LDL-C) and high-density lipoprotein cholesterol (HDL-C) is important. In this study, we aimed to investigate the effect of remnant cholesterol on LDL-C and HDL-C levels measured using homogenous methods.

METHODS

We obtained 41 commutable frozen serum samples and measured LDL-C and HDL-C levels. Eight measurement systems were used, and the degree of bias was obtained by comparing with the values obtained using the reference methods. Correlations among remnant cholesterol/LDL-C, remnant cholesterol/HDL-C, and bias were analyzed using Spearman's analysis.

RESULTS

In all eight systems, samples with a positive bias >4 % had lower LDL-C levels and higher remnant cholesterol levels, as measured by the reference methods, compared to those with a bias ≤4 %. A significant correlation between remnant cholesterol/LDL-C and a positive bias of LDL-C was observed in six of the eight systems evaluated. For HDL-C bias, three systems showed a positive correlation, and three systems showed a negative correlation. In some systems, LDL-C bias was higher in samples with remnant cholesterol/LDL-C ≥0.25 than in those with remnant cholesterol/LDL-C <0.25.

CONCLUSIONS

Remnant cholesterol has a potential effect on direct LDL-C and HDL-C measurements, which has been observed when several measurement systems are used. For these systems, manufacturers should improve the methods to reduce the interference of remnant cholesterol.

摘要

目的

识别直接测量低密度脂蛋白胆固醇(LDL-C)和高密度脂蛋白胆固醇(HDL-C)时潜在的偏差来源非常重要。在本研究中,我们旨在探讨残留胆固醇对采用均相方法测量的LDL-C和HDL-C水平的影响。

方法

我们获取了41份可交换的冷冻血清样本,并测量了LDL-C和HDL-C水平。使用了8种测量系统,通过与参考方法获得的值进行比较得出偏差程度。使用Spearman分析来分析残留胆固醇/LDL-C、残留胆固醇/HDL-C与偏差之间的相关性。

结果

在所有8种系统中,与偏差≤4%的样本相比,偏差>4%的样本通过参考方法测量时具有更低的LDL-C水平和更高的残留胆固醇水平。在评估的8种系统中的6种中,观察到残留胆固醇/LDL-C与LDL-C的正向偏差之间存在显著相关性。对于HDL-C偏差,3种系统呈正相关,3种系统呈负相关。在一些系统中,残留胆固醇/LDL-C≥0.25的样本中的LDL-C偏差高于残留胆固醇/LDL-C<0.25的样本。

结论

残留胆固醇对直接LDL-C和HDL-C测量有潜在影响,在使用多种测量系统时已观察到这一点。对于这些系统,制造商应改进方法以减少残留胆固醇的干扰。