Home Tonometry Diurnal Intraocular Pressure Patterns, Patient Adherence, and Measurement Reliability in a Prospective Clinical Cohort.

作者信息

Perkins Scott W, Joo Julia H, Allan Kevin C, Sood Shalini, Schulgit Matthew James, Markakis George A, Li Ang

机构信息

Cleveland Clinic Lerner College of Medicine, Case Western Reserve University, Cleveland, OH, USA.

Cole Eye Institute, Cleveland Clinic, Cleveland, OH, USA.

出版信息

Clin Ophthalmol. 2025 Sep 26;19:3547-3556. doi: 10.2147/OPTH.S545165. eCollection 2025.

Abstract

PURPOSE

Assess patient adherence, diurnal intraocular pressure patterns, and correlation with in-office tonometry of home intraocular pressure monitoring (IOP).

PATIENTS AND METHODS

In this prospective cohort study, 75 patients with open-angle disease (150 eyes) were recruited and loaned the iCare HOME2 tonometer (ICH2) after in-person training and instructed to take 6 measurements per day for 10 ± 4 days. Statistical analysis of adherence, diurnal IOP patterns, and correlation with in-office tonometry was conducted.

RESULTS

Most patients had good adherence (65% of patients took ≥ 6 measurements per day, median 7.4 measurements/day). However, adherence was poor for a notable minority of patients (19% of patients took less than 2 measurements per day). Patients with low adherence reported difficulty obtaining measurements and desire for more training with the ICH2 device. ICH2 identified diurnal patterns of IOP fluctuation: IOP varied significantly over the 24-hour day (p < 0.0001), and early morning peak IOP at 3 AM decreased significantly throughout the day to minimum mean IOP at 10 PM (p < 0.0001). IOP variance was significantly greater outside of clinic hours (p < 0.0001) and 36% of patients had maximum IOP outside of clinic hours. In-office IOP measurements had significant positive correlation to the closest ICH2 measurement obtained by the patient in time for both Goldman applanation (r = 0.35, p < 0.0001) and iCare IC100 (r = 0.48, p < 0.0001).

CONCLUSION

ICH2 is clinically useable and effective for detecting out-of-office IOP spikes, diurnal IOP patterns, and IOP variation.

摘要

目的

评估患者的依从性、眼压的昼夜模式以及与门诊眼压测量的家庭眼压监测(IOP)之间的相关性。

患者与方法

在这项前瞻性队列研究中,招募了75例开角型青光眼患者(150只眼),他们在接受现场培训后借用iCare HOME2眼压计(ICH2),并被要求每天测量6次,持续10±4天。对依从性、眼压昼夜模式以及与门诊眼压测量的相关性进行了统计分析。

结果

大多数患者依从性良好(65%的患者每天测量≥6次,中位数为7.4次/天)。然而,相当一部分患者依从性较差(19%的患者每天测量少于2次)。依从性低的患者表示难以进行测量,并希望对ICH2设备进行更多培训。ICH2识别出眼压波动的昼夜模式:眼压在24小时内有显著变化(p<0.0001),凌晨3点的清晨眼压峰值在一天中显著下降,至晚上10点降至最低平均眼压(p<0.0001)。非门诊时间的眼压方差显著更大(p<0.0001),36%的患者在非门诊时间眼压最高。门诊眼压测量与患者及时获得的最接近的ICH2测量值在Goldman压平眼压计(r=0.35,p<0.0001)和iCare IC100(r=0.48,p<0.0001)方面均有显著正相关。

结论

ICH2在临床上可用于检测门诊外眼压峰值、眼压昼夜模式和眼压变化,且效果良好。