Young H W, Larson E W, Dominik J W
Appl Microbiol. 1974 Dec;28(6):929-34. doi: 10.1128/am.28.6.929-934.1974.
The May spinning top generator was adapted to a modified Henderson tube for producing large aerosol particles (>4 mum) to obtain almost exclusive upper respiratory tract deposition of infectious aerosols in exposed mice. The system was installed in a biological safety cabinet to permit experimentation with pathogens. A novel mechanism utilizing parts from a machinists micrometer and the mechanical stage from a light microscope was developed for the spinning top generator as a means for precisely positioning the liquid feed needle. Aerosol light-scatter properties were continuously analyzed to provide relative measures of particle size distribution and aerosol concentration. When mice were exposed to influenza virus aerosols in which none of the virus was contained in particles with aerodynamic diameters <4 mum, essentially all of the virus was deposited in the upper respiratory tract tissues.
五月旋转顶部发生器被改装成一个改良的亨德森管,用于产生大的气溶胶颗粒(>4微米),以便在暴露的小鼠中使感染性气溶胶几乎完全沉积在上呼吸道。该系统安装在生物安全柜中,以允许对病原体进行实验。为旋转顶部发生器开发了一种利用机械师千分尺部件和光学显微镜机械载物台的新颖机制,作为精确放置液体进料针的一种手段。对气溶胶的光散射特性进行持续分析,以提供颗粒大小分布和气溶胶浓度的相对测量值。当小鼠暴露于空气动力学直径<4微米的颗粒中不含任何病毒的流感病毒气溶胶时,基本上所有病毒都沉积在上呼吸道组织中。