Improved accuracy of continuous measurement of arterial oxygen tension in sick newborn infants. Criteria for reading and recalibrating the electrode.

作者信息

Wilkinson A R, Phibbs R H, Gregory G A

出版信息

Arch Dis Child. 1979 Apr;54(4):307-10. doi: 10.1136/adc.54.4.307.

Abstract

Improved accuracy of continuous arterial oxygen tension (PaO2) measurement is obtained if allowance is made for the response time of the electrode. Recalibration is necessary only when PaO2 measured in a blood sample differs from the electrode value by more than +/- 11.5 mmHg (1.5 kPa).

摘要

如果考虑到电极的响应时间,连续动脉血氧分压(PaO2)测量的准确性会得到提高。仅当血液样本中测得的PaO2与电极值相差超过±11.5 mmHg(1.5 kPa)时,才需要重新校准。