Abstract

One hundred and twenty five measurements of arterial oxygen saturation (Stcao2) obtained with a transcutaneous pulse oximeter were compared with direct arterial oxygen tension (Pao2) in 13 preterm infants with predominantly fetal haemoglobin. Stcao2 of 86-92% corresponded to Pao2 of 5-13 kPa. Stcao2 above 92%, however, was sometimes associated with Pao2 above 13 kPa.

摘要

对13名主要为胎儿血红蛋白的早产儿,使用经皮脉搏血氧仪测量了125次动脉血氧饱和度(Stcao2),并与直接动脉血氧分压(Pao2)进行了比较。Stcao2为86%-92%时,对应的Pao2为5-13千帕。然而,Stcao2高于92%时,有时会伴有Pao2高于13千帕。

相似文献

1
Pulse oximetry in preterm infants.早产儿的脉搏血氧饱和度测定
Arch Dis Child. 1987 Sep;62(9):957-8. doi: 10.1136/adc.62.9.957.
8
Targeting Oxygen in Term and Preterm Infants Starting at Birth.针对出生时的足月和早产儿的氧靶向治疗。
Clin Perinatol. 2019 Sep;46(3):459-473. doi: 10.1016/j.clp.2019.05.013. Epub 2019 Jun 12.

引用本文的文献

4
High or low oxygen saturation for the preterm baby.早产儿的高氧或低氧饱和度
Arch Dis Child Fetal Neonatal Ed. 2001 May;84(3):F149. doi: 10.1136/fn.84.3.F149.
6
Uses and abuses of pulse oximetry.脉搏血氧饱和度测定法的应用与滥用
Arch Dis Child. 1996 Jan;74(1):77-80. doi: 10.1136/adc.74.1.77.
7
The pulse oximeter.脉搏血氧仪。
Indian J Pediatr. 1993 May-Jun;60(3):469-70. doi: 10.1007/BF02751220.
10
Pulse oximetry.脉搏血氧饱和度测定法
J Clin Monit. 1989 Jan;5(1):37-62. doi: 10.1007/BF01618369.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验