Continuous monitoring of arterial oxygen tension using a catheter-tip polarographic electrode in infants.

作者信息

Conway M, Durbin G M, Ingram D, McIntosh N, Parker D, Reynolds E O, Soutter L P

出版信息

Pediatrics. 1976 Feb;57(2):244-50.

PMID:1250658
Abstract

An oxygen electrode mounted in the tip of an umbilical artery catheter was used in 36 newborn infants with severe respiratory illnesses, 28 of whom survived. Thirty-seven electrodes were used. The median age at insertion was 4 hours (range, 30 minutes to 122 hours). Three electrodes failed to work and they were removed or replaced, and two could not be properly evaluated. Thirty-two electrodes functioned satisfactorily for 10 to 190 hours (mean, 75 hours) after a one-point calibration against blood sampled through the catheter. Twenty-two did not need recalibrating before they were removed after 10 to 190 hours (mean, 88 hours. Four of the remaining ten electrodes were recalibrated once after 33 to 97 hours and then functioned until removed 15 to 55 hours later. The other six electrodes failed after 32 to 105 hours (mean, 49 hours). Complications were few. A total of 356 arterial blood samples, obtained after the initial calibration and before any recalibration was necessary, gave a correlation coefficient of 0.93 (P less than .0001) against an independent system for measuring arterial oxygen tension (Pao2) (Radiometer Type E.5046 oxygen electrode). We conclude that the catheter-tip electrode is a safe and reliable instrument for continuously recording Pao2 in newborn infants which much simplifies the management of serious respiratory illnesses.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验