Lappin T R, Lamont A, Nelson M G
J Clin Pathol. 1972 Jun;25(6):539-42. doi: 10.1136/jcp.25.6.539.
For counting erythrocytes the instrument was precise, with a mean coefficient of variation of 1.21%. Erythrocyte counts showed close agreement with results obtained on a Coulter A electronic counter of proven accuracy. When the Celloscope 401 was modified by the manufacturers to eliminate electrical interference from other laboratory equipment, satisfactory precision and accuracy for white cell counting was obtained. Using cetrimide diluent the coefficient of variation was 1.6% but when using saponin/saline diluent the coefficient of variation was 3.5%. For leucocyte counting there was close agreement between duplicate tests performed on the Celloscope 401 and the Coulter S. The instrument was capable of satisfactory precision and accuracy in platelet counting, provided that the sedimentation method was used to obtain a platelet-rich plasma. The best results were obtained if a two-step dilution was carried out with a first dilution in 10% EDTA and a second in 2.5 mM cocaine in water. Using this method the precision study indicated a coefficient of variation of 3.11%. Close agreement was obtained between platelet counts on the Celloscope 401 when compared with the results obtained either by phase-contrast microscopy or using another electronic counter. Allowing for predilution and duplicate counts on each sample, the rate of throughput was approximately 32 samples per hour. Throughout the test period, the instrument remained electronically and mechanically stable.
用于红细胞计数时,该仪器很精确,平均变异系数为1.21%。红细胞计数结果与用经证实准确性的库尔特A电子计数器获得的结果高度一致。当制造商对Celloscope 401进行改进以消除其他实验室设备的电气干扰后,白细胞计数获得了令人满意的精密度和准确度。使用溴化十六烷基三甲铵稀释液时变异系数为1.6%,但使用皂素/盐水稀释液时变异系数为3.5%。对于白细胞计数,在Celloscope 401和库尔特S上进行的重复测试结果高度一致。如果采用沉降法获得富含血小板的血浆,该仪器在血小板计数方面能够达到令人满意的精密度和准确度。如果进行两步稀释,第一步用10%乙二胺四乙酸(EDTA)稀释,第二步用2.5 mM可卡因水溶液稀释,可获得最佳结果。采用这种方法进行精密度研究,变异系数为3.11%。将Celloscope 401上的血小板计数结果与相差显微镜检查或使用另一台电子计数器获得的结果相比,两者高度一致。考虑到每个样本的预稀释和重复计数,通量约为每小时32个样本。在整个测试期间,该仪器在电子和机械方面均保持稳定。