An evaluation of the autoanalyzer SMA-4.

作者信息

Lappin T R, Lamont A, Nelson M G

出版信息

J Clin Pathol. 1969 Jan;22(1):11-8. doi: 10.1136/jcp.22.1.11.

Abstract

The Technicon AutoAnalyzer SMA-4 has been systematically evaluated. Carry-over from the sample cups was found to be within an acceptable range. The precision and accuracy of the four parameters determined by the instrument have been investigated. The haemoglobin results were found to be accurate. There was good agreement with leucocyte counts performed on the model A Coulter electronic cell counter. After certain modifications had been made in the manifold, satisfactory degrees of accuracy were also obtained for the erythrocyte counts. Although the determination of the haematocrit by conductance is influenced by a variety of factors, the mean coefficient of variation was found to be 0.83% and 95% of the results agreed within +/- 2% of those obtained by the microhaematocrit centrifugation method.A logistic assessment of the SMA-4 when it was put into routine use indicated the need to select blood samples without excess anticoagulant, to calibrate the instrument before each run, to test for instrumental drift, to use a rapid method of correcting stoppages, and, if concurrent reporting is carried out, to employ the part-time services of a second operator. The presence of dust in the environment was found to have a deleterious effect on both the mechanical and electronic components of the system requiring preventive maintenance. It should be possible, by interfacing an analogue-digital converter and automatic punch, to produce data compatible with records.

摘要

相似文献

1
An evaluation of the autoanalyzer SMA-4.
J Clin Pathol. 1969 Jan;22(1):11-8. doi: 10.1136/jcp.22.1.11.
2
Multichannel continuous flow analysis on the SMA-4--7A.在SMA - 4 - 7A上进行多通道连续流动分析。
J Clin Pathol Suppl Coll Pathol. 1969;3:20-5. doi: 10.1136/jcp.s2-3.1.20.
4
An evaluation of the Coulter model S.
J Clin Pathol Suppl Coll Pathol. 1969;3:26-33. doi: 10.1136/jcp.s2-3.1.26.
5
An assessment of the Coulter counter model S.库尔特计数器S型的评估
J Clin Pathol. 1970 Feb;23(1):68-76. doi: 10.1136/jcp.23.1.68.
6
Data processing in haematology.血液学中的数据处理。
J Clin Pathol. 1972 Jul;25(7):594-8. doi: 10.1136/jcp.25.7.594.
8
An evaluation of the Haema-Count MK-40 blood counting system.
Am J Clin Pathol. 1976 Nov;66(5):883-91. doi: 10.1093/ajcp/66.5.883.
10

引用本文的文献

1
An assessment of the Coulter counter model S.库尔特计数器S型的评估
J Clin Pathol. 1970 Feb;23(1):68-76. doi: 10.1136/jcp.23.1.68.
2
An evaluation of the Fisher Hem-Alyzer.费希尔血液分析仪的评估。
J Clin Pathol. 1970 Feb;23(1):60-4. doi: 10.1136/jcp.23.1.60.
3
Multichannel continuous flow analysis on the SMA-4--7A.在SMA - 4 - 7A上进行多通道连续流动分析。
J Clin Pathol Suppl Coll Pathol. 1969;3:20-5. doi: 10.1136/jcp.s2-3.1.20.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验