Flachenecker G, Fastenmeier K
J Urol. 1979 Sep;122(3):336-41. doi: 10.1016/s0022-5347(17)56395-8.
Transurethral resection has been simulated electrically to detect the current pathways and locations of heat production inside the patient. Different electrical behavior was noted with instruments using electrically insulated or metal sheaths. Conductive lubricating jelly should be used with the metal sheath and non-conducting jelly should be used with the insulated sheath.
已经通过电模拟经尿道切除术来检测患者体内的电流路径和产热位置。使用电绝缘或金属鞘的器械观察到了不同的电行为。金属鞘应使用导电润滑凝胶,绝缘鞘应使用非导电凝胶。