股骨颈骨折内固定的实验研究

Experimental study on internal fixation of femoral neck fractures.

作者信息

Van Audekercke R, Martens M, Mulier J C, Stuyck J

出版信息

Clin Orthop Relat Res. 1979 Jun(141):203-12.

PMID:477109
Abstract

The relative strength and the failure mode of internal fixation by multiple pinning and nailing was investigated on 35 specimens. Internal fixation consisted of 3, 5 and 7 Knowles pins, a Smith-Petersen nail or a sliding nail plate. The strength of a specimen with a triflanged nail is only half of that with multiple Knowles pins. There is no significant difference in strength of 3, 5, or 7 pin specimens. The strength of an internal fixation with a sliding nail plate is not superior to the strength of multiple pinning. The mode of failure for the different types of internal fixation is primarily a crush fracture of the supporting trabecular bone at the femoral neck with downward and outward migration of the pins or nail. Except with the fixation with 3 pins where an initial bending could be observed, failure is not a plastic deformation of the internal fixation device.

摘要

对35个标本研究了多针和钉内固定的相对强度及失效模式。内固定包括3枚、5枚和7枚诺尔斯针、一枚史密斯-彼得森钉或一枚滑动钉板。使用三叶钉的标本强度仅为使用多枚诺尔斯针标本强度的一半。3枚、5枚或7枚针标本的强度无显著差异。滑动钉板内固定的强度并不优于多针固定。不同类型内固定的失效模式主要是股骨颈支撑小梁骨的挤压骨折,伴有针或钉向下和向外移位。除了3枚针固定时可观察到初始弯曲外,失效并非内固定装置的塑性变形。