Esakowitz L, Kriss A, Shawkat F
Eye Department, Hospital for Sick Children, London, UK.
Eye (Lond). 1993;7 ( Pt 1):169-71. doi: 10.1038/eye.1993.36.
Single flash scotopic and photopic electroretinograms (ERGs) were recorded from the same subjects using six types of corneal electrode, in order to assess their relative effectiveness. In addition, the ERG from a lower eyelid skin electrode was recorded to give an indication of the degree of attenuation to be expected from a skin electrode. On average, the scotopic ERG recorded from the Burian Allen electrode measured 471 microV (100%), and relative to this the b-waves recorded using other electrodes were as follows: JET (89%), C-glide (77%), gold foil (56%), DTL (46%) and skin (12%). Under photopic conditions the order was the same and the interelectrode proportions similar. The b-wave amplitude recorded using the Burian Allen electrode was 125 microV (100%), and with other electrodes was as follows: JET (93%), C-glide (78%), gold foil (60%), DTL (60%) and skin (14%).
为了评估六种角膜电极的相对有效性,对同一受试者记录了单闪光暗视和明视视网膜电图(ERG)。此外,还记录了来自下眼睑皮肤电极的ERG,以表明皮肤电极预期的衰减程度。平均而言,从Burian Allen电极记录的暗视ERG测量值为471微伏(100%),相对于此,使用其他电极记录的b波如下:JET(89%)、C-glide(77%)、金箔(56%)、DTL(46%)和皮肤(12%)。在明视条件下,顺序相同,电极间比例相似。使用Burian Allen电极记录的b波振幅为125微伏(100%),使用其他电极的情况如下:JET(93%)、C-glide(78%)、金箔(60%)、DTL(60%)和皮肤(14%)。