Steinmetz J, Panek E, Gaspart E
J Clin Chem Clin Biochem. 1979 Aug;17(8):547-51. doi: 10.1515/cclm.1979.17.8.547.
Cholesterol was assayed on the Greiner GSA II and the Technicon SMA 12/60 and SMA II autoanalyzers, with Allain's entirely enzymic method (Allain et al. (1974), Clin. Chem. 20, 470--475). The coefficients of variation in the results over a period of one month were less than 6% with the GSA II and the SMA II, but up to 9% on the SMA 12/60. Although the accuracy of the enzymic method depends on the specificity of the cholesterol oxidase and cholesterol esterase, the results were in good agreement with those obtained by Abell's method (Abell et al. (1952), J. Biol. Chem. 195, 357--366) on fresh or freeze dried human plasma. The cholesterol concentration was stable at 4 degrees C, or in frozen or freeze dried plasma. None of the drugs tested interfered with the enzymic assay.
采用Greiner GSA II和Technicon SMA 12/60以及SMA II自动分析仪,运用阿兰全酶法(Allain等人,(1974),《临床化学》20,470 - 475)对胆固醇进行测定。在一个月的时间里,GSA II和SMA II的结果变异系数小于6%,但SMA 12/60的变异系数高达9%。尽管酶法的准确性取决于胆固醇氧化酶和胆固醇酯酶的特异性,但这些结果与用阿贝尔法(Abell等人,(1952),《生物化学杂志》195,357 - 366)在新鲜或冻干人血浆中获得的结果高度一致。胆固醇浓度在4℃、冷冻或冻干血浆中均保持稳定。所测试的药物均未干扰酶法测定。