Brewster G M, Evans A L
Br Heart J. 1979 Sep;42(3):266-70. doi: 10.1136/hrt.42.3.266.
Displacement of endocardial pacemaker leads from the initial site at the time of implantation is a major cause of unpredictable pacemaker failure. In this survey, we have made a retrospective study of the effectiveness of the attempts made by two manufacturers to design leads to reduce the frequency of this occurrence. The cumulative displacement records of 5 types of leads (Medtronic types 5818, 6904, 6901, and 6950, and Cardiac Pacemakers Inc. type 4210) have been studied, and show that Medtronic 5818 and 6950, and CPI 4210 leads are superior in this respect.
植入时心内膜起搏器导线从初始位置移位是起搏器意外失效的主要原因。在本次调查中,我们对两家制造商为设计导线以降低这种情况发生频率所做尝试的有效性进行了回顾性研究。已对5种类型导线(美敦力5818型、6904型、6901型和6950型,以及心脏起搏器公司4210型)的累积移位记录进行了研究,结果表明美敦力5818型和6950型以及CPI 4210型导线在这方面表现更优。