Suppr超能文献

正常入射和斜入射钴-60远距离治疗束的剂量测定

Dosimetry of normally and obliquely incident cobalt-60 teletherapy beams.

作者信息

Sabuwala H A

出版信息

Strahlentherapie. 1979 Nov;155(11):766-9.

PMID:516098
Abstract

Standard isodose charts are generally given by the manufacturer for normal incidence of radiation beams of various qualities. These charts are used for routine treatment planning in a radiotherapy department. Due to the enormous variety of the cases encountered in the department, the physicist often runs short of appropriate charts required for specific treatment planning. Besides, the plan sometimes needs a beam directed at an angle to the patient's surface. As a result, isodose charts for individual treatment plans are desirable for precise delivery of the requisite dose to the tumour volume. In this paper a computer method is described for determining dose distributions for normally and obliquely incident 60Co beam. The experimental verification of the computed results using a 0.4 cm3 ionization chamber and a perspex water phantom assembly is also presented. The results are found to be in good agreement.

摘要

标准等剂量曲线通常由制造商给出,用于各种质量的辐射束的正常入射情况。这些曲线用于放射治疗科的常规治疗计划。由于科室中遇到的病例种类繁多,物理学家常常缺乏特定治疗计划所需的合适曲线。此外,治疗计划有时需要一束与患者体表成一定角度的射束。因此,对于将所需剂量精确地传递到肿瘤体积,单个治疗计划的等剂量曲线是很有必要的。本文描述了一种计算机方法,用于确定60Co射束正常入射和斜入射时的剂量分布。还给出了使用0.4 cm3电离室和有机玻璃水模体组件对计算结果进行的实验验证。结果发现两者吻合良好。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验