Leunissen J L, Pluygers H W, Elbers P F
J Microsc. 1979 Dec;117(3):355-62. doi: 10.1111/j.1365-2818.1979.tb04692.x.
Modifications of the Reichert FC-2 cryomicrotome attachment which facilitate sectioning of frozen material are described. Most valuable was the construction of a specimen holder chuck made of brass and connected to the specimen arm with a glass rod. Other changes include improved shielding from heat influx, a constant nitrogen pressure cooling system and a stable cryochamber mount. These modifications considerably improve reproducibility.