Ophthalmology. 1979 Aug;86(8):1510-2. doi: 10.1016/s0161-6420(79)35369-6.
In a prospective evaluation of 39 patients who had arteriography for suspected carotid stenosis, suction ophthalmodynamometry correctly identified 86% of the patients who had radiographically demonstrable unilateral or asymmetric obstructions. This compares favorably with carotid compression tonography, oculoplethysmography, and Doppler flow studies. The false-positive rate was 27%, a figure higher than the latter two noninvasive tests.