Pinkerton P H, Spence I, Ogilvie J C, Ronald W A, Marchant P, Ray P K
J Clin Pathol. 1970 Feb;23(1):68-76. doi: 10.1136/jcp.23.1.68.
An assessment of the Coulter model S automatic blood counter has been carried out. The standard deviations for the haemoglobin concentration, haematocrit, red cell count, and white cell count are, respectively, +/- 0.15 g/100 ml, +/- 0.45%, +/- 0.04 m/c mm, and +/- 0.47 thous/c mm. These results are clearly more accurate than careful manual estimates, performed for comparison on the same samples. Details of the comparisons are presented. A comparison is also made with routine daily estimations.;Carryover' from one sample to the next was found to be about 2%, and tests in the ranges likely to be found in practice showed good linearity for the haemoglobin estimation, haematocrit, red cell count, and white cell count. A brief account of instrument failures is given.The performance of the Coulter model S compares well with that of other automated equipment for which detailed evaluations are available.
已对库尔特S型自动血液计数器进行了评估。血红蛋白浓度、血细胞比容、红细胞计数和白细胞计数的标准差分别为±0.15 g/100 ml、±0.45%、±0.04百万/mm和±0.47千/mm。这些结果显然比为进行比较而对相同样本进行的仔细手工估算更为准确。文中给出了比较的详细情况。还与日常常规估算进行了比较。发现从一个样本到下一个样本的“残留”约为2%,在实际可能出现的范围内进行的测试表明,血红蛋白估算、血细胞比容、红细胞计数和白细胞计数具有良好的线性。文中简要介绍了仪器故障情况。库尔特S型的性能与其他有详细评估的自动化设备相比表现良好。