Suppr超能文献

用于原位非接触式记录暴露动脉直径的光电装置。

Photoelectric device for contact-free recording of the diameters of exposed arteries in situ.

作者信息

Wetterer E, Busse R, Bauer R D, Schabert A, Summa Y

出版信息

Pflugers Arch. 1977 Mar 11;368(1-2):149-52. doi: 10.1007/BF01063468.

Abstract

A photoelctric device is described which permits contact-free recording of the diameter of an in situ exposed artery. The light emitted by the filament of a bulb (6 W) is collimated by a projection lens (1:2.8, 50 mm) and directed onto the surface of a silicone photocell (12.6X6.2 mm) covered by a neutral light filter. The voltage drop caused by the photocell current flowing through an adjustable load resistor represents the signal voltage which is fed into the input of an instrumentation amplifer. The artery under investigation is positioned above the photocell parallel to its short axis. The signal voltage decreases linearly with increasing area of the shadow cast by the artery on the photocell and thus the output voltage of the amplifier is linearly proportional to the diameter of the artery. The combined frequency response of photocell and amplifier was examined by sinusoidally modulated light emitted by a luminescent diode. The amplitude ratio was constant up to 200 Hz and the time lag was about 0.2 ms. The device was used on the canine carotid and femoral arteries. An example is shown.

摘要

描述了一种光电装置,它可以对原位暴露动脉的直径进行非接触式记录。灯泡(6瓦)灯丝发出的光由投影透镜(1:2.8,50毫米)准直,并照射到覆盖有中性滤光片的硅光电管(12.6×6.2毫米)表面。由流经可调负载电阻的光电池电流引起的电压降代表信号电压,该信号电压被馈入仪表放大器的输入端。被研究的动脉平行于其短轴放置在光电管上方。信号电压随着动脉在光电管上投射的阴影面积增加而线性下降,因此放大器的输出电压与动脉直径成正比。通过发光二极管发出的正弦调制光来检测光电管和放大器的组合频率响应。在高达200赫兹时,幅度比是恒定的,时间滞后约为0.2毫秒。该装置用于犬的颈动脉和股动脉。给出了一个示例。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验