Chalon J, Simon R, Patel C, Ramanathan S, Sessler S, Turndorf H
Anesth Analg. 1978 May-Jun;57(3):307-12.
A pediatric anesthesia system was constructed by connecting the machine end of a modified Mapleson D system to a soda lime canister containing a water vaporizer heated through the reaction of neutralization of the lime by the CO2 emerging from the breathing tube. The vaporizer was placed in line with the fresh-gas delivery tube of the circuit. The humidity output of the system varied from 15 +/- 1 mg H2O/L with a CO2 output of 15 ml/L to 27 +/- 1.5 mg H2O/L with a CO2 output of 60 ml/min. Rebreathing of CO2 was within acceptable limits as long as fresh-gas inflow remained at 5 L/min. The system compares favorably in respect to its humidity output with all other circuits availabe on the market for pediatric anesthesia.
通过将改良的Mapleson D系统的机器端连接到一个装有水汽化器的苏打石灰罐构建了一个小儿麻醉系统,该水汽化器通过呼吸管中排出的二氧化碳与石灰的中和反应来加热。汽化器与回路的新鲜气体输送管串联放置。该系统的湿度输出在二氧化碳输出为15 ml/L时为15±1 mg H₂O/L,在二氧化碳输出为60 ml/min时为27±1.5 mg H₂O/L。只要新鲜气体流入量保持在5 L/min,二氧化碳的再吸入就在可接受的限度内。就湿度输出而言,该系统与市场上所有其他可用的小儿麻醉回路相比具有优势。