Suppr超能文献

使用微型电极在体内测量二氧化碳张力。

In vivo measurement of carbon dioxide tension with a miniature electrode.

作者信息

Olievier C N, Berkenbosch A, Quanjer P H

出版信息

Pflugers Arch. 1978 Mar 20;373(3):269-72. doi: 10.1007/BF00580834.

Abstract

A commercially available catheter type electrode with which PCO2 can be continuously measured in vivo and in vitro gave progressively less accurate results the longer the measuring period was extended. This proved to be due to temperature effects and a change in sensitivity with time. A correction procedure for these effects was developed which was based on two observations. 1. The relationship between temperature and the logarithm of the sensitivity of the electrode-amplifier combination was linear and virtually identical for 9 electrodes: 8% change in sensitivity for a deviation of 1 degree C from the temperature during calibration. 2. The change in sensitivity due to drift of the electrode output is approximately a logarithmic function of time: 1 h after calibration all electrodes exhibited a decreased sensitivity, varying between 0.3 and 16.7%. The drift effect can be dealt with by repeated calibrations, preferably at 1 1/2 h intervals. The adequacy of the correction procedure was assessed in in vivo measurements in cats and dogs. The mean PCO2 difference between the in vivo measurement, corrected for temperature and drift, and samples analyzed with a conventional electrode, was 0.005 kPa (0.04 mm Hg) with a standard deviation of 0.187 kPa (1.39 mm Hg).

摘要

一种可在体内和体外连续测量PCO2的市售导管式电极,测量时间越长,其结果的准确性逐渐降低。事实证明,这是由于温度效应和灵敏度随时间的变化所致。基于两项观察结果,开发了针对这些效应的校正程序。1. 温度与电极-放大器组合灵敏度的对数之间的关系呈线性,9个电极的情况几乎相同:在校准期间,温度偏离1摄氏度时,灵敏度变化8%。2. 由于电极输出漂移导致的灵敏度变化大致是时间的对数函数:校准后1小时,所有电极的灵敏度均下降,下降幅度在0.3%至16.7%之间。漂移效应可通过重复校准来处理,最好每隔1.5小时校准一次。在校猫和狗的体内测量中评估了校正程序的充分性。经温度和漂移校正后的体内测量值与用传统电极分析的样本之间的平均PCO2差值为0.005 kPa(0.04 mmHg),标准差为0.187 kPa(1.39 mmHg)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验