Evaluation of a dual-function pH and PCO2 in vivo sensor.

作者信息

Coon R L, Lai N C, Kampine J P

出版信息

J Appl Physiol. 1976 Apr;40(4):625-9. doi: 10.1152/jappl.1976.40.4.625.

Abstract

A newly developed, dual-function pH and PCO2 sensor was evaluated in this study. The sensors were placed in the femoral arteries of dogs anesthetized with sodium pentobarbital. Comparisons were made between systemic arterial pH and PCO2 measured using the sensor and those measured from blood samples drawn at 15-min intervals over a 7-h period using a bench instrument. The mean pH of the bench instrument measurements was 7.43. The mean difference of the sensor measurements from the bench instrument measurements for 207 comparisons was 0.0003 pH +/- 0.061 SD. The mean PCO2 of the bench instrument measurements was 40 mmHg. The mean difference of the sensor measurements from those of the bench instrument for 212 comparisons was -1.43 mmHg +/- 5.17 SD. The sensors performed equally well in the presence of metabolic or respiratory acidosis and alkalosis. The dual-function sensors evaluated in this study are useful for trend monitoring of pH and PCO2 over at least a 7-h period without recalibration. With improvement in the consistency of sensor construction, these sensors will be reliable in vivo sensing devices for blood pH and PCO2 and thus valuable research and clinical instruments.