Regan D
Electroencephalogr Clin Neurophysiol. 1976 Jun;40(6):654-60. doi: 10.1016/0013-4694(76)90140-1.
This note describes a method for rapidly measuring the latency of a steady-state visual evoked potential (EP). Three sinewaves of different frequencies (F1, F2 and F3 c/sec) are summed, and the resulting waveform modulates either light intensity (for flicker stimulation) or spatial contrast (for pattern stimulation). Three evoked potentials of frequencies F1, F2 and F3 c/sec respectively are stimultaneously recorded by running three Fourier analysers in parallel. The outputs of the three analysers are presented in a single polar plot so as directly to display the three values of EP amplitude and phase. Practical methods for generating the pattern and flicker stimuli are described.