End-diastolic and end-systolic ventricular volume clamper for isolated canine heart.

作者信息

Suga H, Sagawa K

出版信息

Am J Physiol. 1977 Dec;233(6):H718-22. doi: 10.1152/ajpheart.1977.233.6.H718.

Abstract

We devised a servo-controlled pump system which can clamp both end-diastolic and end-systolic volumes of the left ventricle of an excised, supported canine heart at desirable values in the face of changing patterns of ejection and filling. The system consists of a Bellofram cylinder and a powerful electromagnetic shaker which is driven by a position servo circuit. A water-filled balloon placed in the ventricle was connected to the water housing of the cylinder. As the ECG triggered the system, a fixed amount of water reciprocated between the heart and the cylinder in a programmed manner. The onset, duration, and speed of ejection and those of filling can be programmed separately. The performance of this pump system was tested in experiments on 15 cross-circulated hearts and proved satisfactory for precise analysis of the instantaneous pressure-volume relation in the ventricle.