Suppr超能文献

活化部分凝血活酶时间的测定。一项国际合作研究。

Measuring partial thromboplastin-time. An international collaborative study.

作者信息

Poller L, Thomson J M

出版信息

Lancet. 1976 Oct 16;2(7990):842-6. doi: 10.1016/s0140-6736(76)91224-1.

Abstract

A series of collaborative exercises on the partial thromboplastin-time (P.T.T.) test, involving over three hundred hospital centres in Britain and overseas, were performed in 1975. Lyophilised test plasmas were issued from the World Health Organisation Collaborating Centre for Anticoagulant Control Reagents to participants, together with a standardised reference P.T.T. reagent and a standard technique. Hospitals were asked to test the plasma samples with the standardised reagent and technique in parallel with their customary local P.T.T. reagent and method. The overall success-rate in detecting the intrinsic clotting abnormality in the eight abnormal test samples was higher with the standardised reagent and technique than with all other reagents. Furthermore, fewer hospitals obtained false positive results when the normal plasma sample was tested with the standardised method rather than with their usual routine reagents. An index was used to measure the success-rate of the P.T.T. reagents in correctly identifying the test plasmas as normal or abnormal. The eight test plasmas showed a varying degree of abnormality. A system of "weighting" was therefore introduced as the failure of a P.T.T. method to detect more severe defects was regarded as more serious. Although hospitals were unfamiliar with the standardised method, the results established its superiority over all other P.T.T. reagents included in the trials in sufficient numbers for analysis. Failures with commercial reagents may have been caused by insensitivity of the cephalin extracts or the unreliability of the manufacturers recommended techniques. Since the same laboratories obtained good results with the standardised method technical failure can be excluded.

摘要

1975年,开展了一系列关于部分凝血活酶时间(PTT)检测的协作试验,涉及英国及海外的三百多家医院中心。世界卫生组织抗凝控制试剂协作中心向参与者发放了冻干测试血浆,同时还提供了标准化参考PTT试剂和标准技术。要求医院使用标准化试剂和技术对血浆样本进行检测,并与当地惯用的PTT试剂和方法同时进行检测。使用标准化试剂和技术检测八个异常测试样本中内源性凝血异常的总体成功率高于所有其他试剂。此外,使用标准化方法检测正常血浆样本时,得出假阳性结果的医院比使用其常规试剂时更少。采用一个指数来衡量PTT试剂正确识别测试血浆为正常或异常的成功率。这八个测试血浆呈现出不同程度的异常。因此引入了一种“加权”系统,因为PTT方法未能检测到更严重的缺陷被认为更为严重。尽管医院对标准化方法不熟悉,但结果表明其优于试验中纳入的所有其他PTT试剂,且数量足以进行分析。商业试剂出现的问题可能是由于脑磷脂提取物不敏感或制造商推荐技术不可靠。由于相同的实验室使用标准化方法取得了良好结果,因此可以排除技术故障。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验