Marlar R A, Bauer P J, Endres-Brooks J L, Montgomery R R, Miller C M, Schanen M M
Am J Clin Pathol. 1984 Oct;82(4):436-9. doi: 10.1093/ajcp/82.4.436.
This study was undertaken to evaluate the precision and sensitivity of three different commercial APTT reagents containing the activators kaolin, micronized silica, or ellagic acid. These reagents varied greatly in their ability to detect mild coagulopathies. The ellagic acid reagent was able to detect the mildest deficiencies for the most common coagulopathies. This reagent was sensitive to 50% levels of Factor VIIIC, whereas the APTT with the kaolin reagent was not prolonged until levels of 35% or less were attained. The micronized silica reagent was the least sensitive to Factor IX deficiency, detecting levels of 12% or less. Precision was similar for all reagents when tested with normal and slightly abnormal plasmas. Since methods and instrumentation vary, each laboratory should evaluate their APTT reagent to determine its precision and sensitivity.
本研究旨在评估三种不同的含有高岭土、微粉硅胶或鞣花酸激活剂的商用活化部分凝血活酶时间(APTT)试剂的精密度和灵敏度。这些试剂在检测轻度凝血障碍的能力上差异很大。鞣花酸试剂能够检测出最常见凝血障碍中最轻微的缺陷。该试剂对Ⅷ因子水平降至50%敏感,而含高岭土试剂的APTT直到Ⅷ因子水平降至35%或更低时才延长。微粉硅胶试剂对Ⅸ因子缺乏最不敏感,检测水平为12%或更低。用正常和轻度异常血浆检测时,所有试剂的精密度相似。由于方法和仪器不同,每个实验室都应评估其APTT试剂以确定其精密度和灵敏度。