起搏器系统分析仪测定脉冲发生器输出电压时的误差来源。

Sources of error in the determination of output voltage of pulse generators by pacemaker system analyzers.

作者信息

Barold S S, Roehrich D R, Falkoff M D, Ong L S, Heinle R A

出版信息

Pacing Clin Electrophysiol. 1980 Sep;3(5):585-96. doi: 10.1111/j.1540-8159.1980.tb05278.x.

Abstract

The use of threshold or pacemaker system analyzers with widely different characteristics has introduced potential sources of error in the determination of the output voltage of pulse generators. This is further compounded by the availability of pulse generators with diverse waveform configurations and programmability capabilities. Because of this non-uniformity, the physician must have some rudimentary knowledge of waveform characteristics and appreciate the limitations of threshold or pacemaker system analyzers to avoid the unnecessary replacement of normally-functioning pulse generators.

摘要

使用具有大不相同特性的阈值或起搏器系统分析仪,在测定脉冲发生器的输出电压时引入了潜在的误差来源。具有不同波形配置和可编程能力的脉冲发生器的可得性使这一情况更加复杂。由于这种不统一性,医生必须具备一些波形特性的基本知识,并了解阈值或起搏器系统分析仪的局限性,以避免不必要地更换功能正常的脉冲发生器。